A Collaborative Study Was Conducted To Determine The Standard Error Of Difference Among Laboratories For Near-Infrared Reflectance Spectroscopic (Nirs) Determination Of Acid-Detergent Fiber (Adf) And Crude Protein In Forages. The 6 Participating Laboratories Were Members Of The Usda/Ars National Near-Infrared Reflectance Spectroscopy Forage Research Project. The Nirs Calibration Equations Were Developed In The Associate Referee's Laboratory For Crude Protein And Adf And Were Transferred To The Instrument In Each Of The Other Collaborating Laboratories. The Calibration Set Included Over 650 Diverse Forage Samples With Crude Protein And Adf Calibration Data; The Validation Set Included 94 Samples Of Bermudagrass. Amonglaboratory Reproducibility For The Nirs Method, Calculated As The Relative Standard Deviation For Reproducibility (Rsdr), Was 1.14% For Adf And 0.42% For Crude Protein. The Variance Component For Among-Laboratory Variation (Coefficient Of Variation) Was 2.54% For Adf And 2.89% For Crude Protein. These Results Confirm That It Is Possible To Calibrate, Validate, And Transfer (Nirs) Equations And Data Among Laboratories For The Accurate Determination Of Adf And Crude Protein, And Thereby Demonstrate That Nirs Can Be Used As A Standard Method For The Analysis Of Forages. The Method Has Been Adopted Official First Action