Abstract
Abstract
The present study evaluated the organ doses, effective doses and conversion factors from the dose–area product to effective dose in pediatric diagnostic cardiac catheterization performed by in-phantom dosimetry and Monte Carlo simulation. The organ and effective doses in 5-y-olds during diagnostic cardiac catheterizations were evaluated using radiophotoluminescence glass dosemeters implanted into a pediatric anthropomorphic phantom and PCXMC software. The mean effective dose was 3.8 mSv (range: 1.8–7.5 mSv). The conversion factors from the dose–area product to effective dose were 0.9 and 1.6 mSv (Gy cm2)−1 for posteroanterior and lateral fluoroscopy, respectively, and 0.9 and 1.5 mSv (Gy cm2)−1 for posteroanterior and lateral cineangiography, respectively. Effective doses evaluated using the pediatric dosimetry system agreed with those obtained using PCXMC software within 12%. The dose data and conversion factors evaluated may guide the estimation of exposure doses in children undergoing diagnostic cardiac catheterization.
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology
Reference21 articles.
1. The ALARA concept in pediatric cardiac catheterization: techniques and tactics for managing radiation dose;Justino;Pediatr. Radiol.,2006
2. Estimation of organ and effective doses for neonate and infant diagnostic cardiac catheterizations;Kawasaki;Am. J. Roentgenol.,2015
3. Estimation of organ doses and effective doses based on in-phantom dosimetry for infant diagnostic cardiac catheterisations with novel X-ray imaging technology;Kawasaki;Radiat. Prot. Dosim.,2018