Abstract
Abstract
In response to the ICRP’s amending the occupational exposure limit for the eye lens, the Institute of Nuclear Energy Research (INER) established the Hp(3) calibration system for eye dose monitoring in Taiwan to accurately assess the dose received in the eye lens. INER employed the narrow-spectrum series radiation according to the ISO 4037 as the X-ray radiation qualities, and the measured half-value layer consistent with a 5% difference. The air kerma rate standard was determined by the self-made free air chamber, and through dose conversion coefficient referring to ISO 4037 to obtain the Hp(3) on an ISO cylinder phantom. Furthermore, the calibration system was provided as the characteristics tests for DOSIRIS headset dosemeters. Finally, the Hp(3) calibration system has been established in Taiwan, and it can be used to provide calibration services for eye lens dosemeters and be applied to the proficiency testing that will be held in 2023.