Affiliation:
1. Radiological Protection Service, La Fe University and Polytechnic Hospital , Valencia , Spain
Abstract
Abstract
The introduction of digital radiography has improved image acquisition. However, rejection of images remains a matter of concern. Reject analysis is part of the quality assurance program in radiology and helps identify potential errors or lack of training. A retrospective study was conducted at the radiology department of a university hospital. The reject rate was calculated both using the number of examinations, $r_n$, and the dose-area product, $r_d$. A reject rate $r_n$ of 3.3% for paediatric units and 4.5% for adults was found. The corresponding values of rd were 4.4 and 8.4%, respectively. The main rejection cause was patient motion, being 50.2% of rejected examinations in adults and 63.7% in children. The contribution of exposure errors was minor, as expected in digital radiography units. A discrepancy between reject rates $r_n$ and $r_d$ was observed, suggesting dosimetric quantities could be considered in reject analysis for further assessment of patient radiation burden.
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology