Simple one-parameter function to retrieve the correct exposure value from CR-39 radon detectors in high-saturation regime. Checks on two kinds of analysis systems

Author:

Marchese Rocco1,Nicolino Antonella2,Palma Achille1,Procopio Salvatore3

Affiliation:

1. ARPAB, Agenzia Regionale per la Protezione dell'Ambiente di Basilicata , 85100 - Potenza, Italy

2. STAR, Southern europe Thomson back-scattering source for Applied Research , 87036 - Rende, Italy

3. ARPACAL, Agenzia Regionale per la Protezione dell'Ambiente della Calabria , 88100 - Catanzaro, Italy

Abstract

Abstract Radon activity concentration measurements in air can be easily performed by using CR-39 detectors whose response is almost linear in the range of medium–low exposures. However, when the exposure values grow too much, a saturation phenomenon becomes relevant and some corrections have to be adopted even if they cannot always be very accurate and easy to apply. Therefore, a simple alternative method for determining the correct response curve of the CR-39 detectors, from low up to very high radon exposures, is showed. To verify its robustness and the general applicability, several certified measurements were carried out in a radon chamber at different levels of exposure. Moreover, two different types of commercially available radon analysis systems were used.

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Reference5 articles.

1. Theoretical calculation on CR-39 response for radon measurements and optimum diffusion chambers dimensions;Askari;Nucl. Inst. Methods Phys. Res. A,2008

2. Study of chemical etching conditions for alpha-particle detection and visualization using solid state nuclear track detectors;Amada;Radiat. Environ. Med.,2019

3. Study of saturation of CR39 nuclear track detectors at high ion fluence and of associated artifact patterns;Gaillard;Rev. Sci. Instrum.,2007

4. Calibration of the Politrack®system based on CR-39 solid-state nuclear track detectors for passive indoor radon concentration measurements;Kropat;Radiat. Prot. Dosim.,2015

5. Funzione semi-empirica ad un parametro per la correzione della saturazione nei rivelatori radon a tracce nucleari CR-39;Marchese;AIRP-Bollettino,2019

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3