In-house 3D-printed aligners: effect of in vivo ageing on mechanical properties

Author:

Can Esad1,Panayi Nearchos2,Polychronis Georgios3,Papageorgiou Spyridon N1ORCID,Zinelis Spiros3,Eliades George3,Eliades Theodore1ORCID

Affiliation:

1. Clinic of Orthodontics and Pediatric Dentistry, Center of Dental Medicine, University of Zurich, Zurich, Switzerland

2. Private Practice, Limassol, Cyprus

3. Department of Biomaterials, School of Dentistry, National and Kapodistrian University of Athens, Athens, Greece

Abstract

Summary Objective To investigate alterations in the mechanical properties of in-house three-dimensional (3D) printed orthodontic aligners after intraoral ageing. Materials and methods Sixteen 3D-printed aligners (TC-85DAC resin, Graphy, Seoul, Korea) were used for the purpose of the study, which were divided into 10 control (not used) aligners and 6 materials retrieved from 4 patients after 1-week service (retrieved group). Samples from the control group were analysed by attenuated total reflectance-Fourier-transform infrared (ATR-FTIR) spectroscopy. Samples from control/retrieved groups were embedded resin and subjected to instrumented indentation testing (IIT) to record force–indentation depth curves, calculating the following (as per ISO 14577-1, 2002 standard): Martens hardness (HM), indentation modulus (EIT), and elastic index (ηIT), and the indentation relaxation index (RIT). Differences between control and retrieved 3D-printed aligners were checked with Mann–Whitney/t-tests at an alpha = 5%. Results ATR-FTIR analysis showed that aligners were made of a vinyl ester–urethane material. The results of the IIT testing were: HM (control: median 91.5 N/mm2, interquartile range [IQR] 88.0–93.0/as-retrieved: median 90.5 N/mm2, IQR 89.0–93.0); EIT (control, mean 2616.3 MPa, standard deviation [SD] 107.0 MPa/retrieved, mean 2673.2 MPa, SD 149.4 MPa); ηIT (control: median 28.6%, IQR 28.2–30.9%/as-retrieved: median 29.0%, IQR 28.7–29.2%); and RIT (control: median 45.5%, IQR 43.0–47.0%/as-retrieved: median 45.1%, IQR 45.0–45.3%). No differences between as-retrieved and control aligners were found for any of the mechanical properties tested (P > 0.05 in all instances). Conclusion The mechanical properties of the in-house 3D-printed aligners tested were not affected after 1 week in service period.

Publisher

Oxford University Press (OUP)

Subject

Orthodontics

Reference25 articles.

1. Attractiveness, acceptability, and value of orthodontic appliances;Rosvall;American Journal of Orthodontics and Dentofacial Orthopedics,2009

2. Patient satisfaction and quality of life status after 2 treatment modalities: Invisalign and conventional fixed appliances;Flores-Mir;American Journal of Orthodontics and Dentofacial Orthopedics,2018

3. Duration of orthodontic treatment with fixed appliances in adolescents and adults: a systematic review with meta-analysis;Abbing;Progress in Orthodontics,2020

4. Treatment outcome with orthodontic aligners and fixed appliances: a systematic review with meta-analyses;Papageorgiou;European Journal of Orthodontics,2020

5. Do it yourself orthodontics: alarmist or real?;Padmanabhan;Journal of Indian Orthodontic Society,2019

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3