Technical note: Assessing gas equilibration systems for continuous <i>p</i>CO<sub>2</sub> measurements in inland waters
-
Published:2016-07-07
Issue:13
Volume:13
Page:3915-3930
-
ISSN:1726-4189
-
Container-title:Biogeosciences
-
language:en
-
Short-container-title:Biogeosciences
Author:
Yoon Tae KyungORCID, Jin Hyojin, Oh Neung-Hwan, Park Ji-Hyung
Abstract
Abstract. High-frequency continuous measurements of the partial pressure of CO2 (pCO2) are crucial for constraining the spatiotemporal dynamics of CO2 emissions from inland water systems. However, direct measurements of pCO2 are scarce, and no systematic comparisons have been conducted on the suitability of the widely used measurement systems for continuous underway or long-term deployment in various field conditions. We compared spray- and marble-type equilibrators and a membrane-enclosed CO2 sensor to assess their suitability for continuous long-term or underway pCO2 measurements in an urbanized river system in Korea. Both equilibrators had a shorter response time compared with the membrane-enclosed sensor, and could capture large spatial variations of pCO2 during a transect study along a highly urbanized river reach. The membrane-enclosed sensor based on passive equilibration provided comparable underway measurements along the river sections where pCO2 varied within the sensor detection range. When deployed in a eutrophic river site, the membrane-enclosed sensor was able to detect large diel variations in pCO2. However, biofouling on the membrane could reduce the accuracy of the measurement during long deployments exceeding several days. The overall results suggest that the fast response of the equilibrator systems facilitates capturing large spatial variations in pCO2 during short underway measurements. However, the attendant technical challenges of these systems, such as clogging and desiccant maintenance, have to be addressed carefully to enable their long-term deployment. The membrane-enclosed sensor would be suitable as an alternative tool for long-term continuous measurements if membrane biofouling could be overcome by appropriate antifouling measures such as copper mesh coverings.
Funder
National Research Foundation of Korea
Publisher
Copernicus GmbH
Subject
Earth-Surface Processes,Ecology, Evolution, Behavior and Systematics
Reference68 articles.
1. Abril, G., Richard, S., and Guérin, F.: In situ measurements of dissolved gases (CO2 and CH4) in a wide range of concentrations in a tropical reservoir using an equilibrator, Sci. Total Environ., 354, 246–251, 2006. 2. Abril, G., Martinez, J.-M., Artigas, L. F., Moreira-Turcq, P., Benedetti, M. F., Vidal, L., Meziane, T., Kim, J.-H., Bernardes, M. C., Savoye, N., Deborde, J., Souza, E. L., Alberic, P., Landim de Souza, M. F., and Roland, F.: Amazon River carbon dioxide outgassing fueled by wetlands, Nature, 505, 395–398, 2014. 3. Abril, G., Bouillon, S., Darchambeau, F., Teodoru, C. R., Marwick, T. R., Tamooh, F., Ochieng Omengo, F., Geeraert, N., Deirmendjian, L., Polsenaere, P., and Borges, A. V.: Technical Note: Large overestimation of pCO2 calculated from pH and alkalinity in acidic, organic-rich freshwaters, Biogeosciences, 12, 67–78, https://doi.org/10.5194/bg-12-67-2015, 2015. 4. Bakker, D. C. E., De Baar, H. J. W., and de Wilde, H. P. J.: Dissolved carbon dioxide in Dutch coastal waters, Mar. Chem., 55, 247–263, 1996. 5. Bastviken, D., Sundgren, I., Natchimuthu, S., Reyier, H., and Gålfalk, M.: Technical Note: Cost-efficient approaches to measure carbon dioxide (CO2) fluxes and concentrations in terrestrial and aquatic environments using mini loggers, Biogeosciences, 12, 3849–3859, https://doi.org/10.5194/bg-12-3849-2015, 2015.
Cited by
29 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|