Total column ozone retrieval from a novel array spectroradiometer

Author:

Egli Luca,Gröbner JulianORCID,Schill Herbert,Maillard Barras ElianeORCID

Abstract

Abstract. This study presents a new total column ozone (TCO) retrieval from the Koherent system, developed at the Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC). The instrument is based on a small, cost-effective, robust, low-maintenance and state-of-the-art-technology array spectroradiometer. It consists of a BTS-2048-UV-S-F array spectroradiometer from Gigahertz-Optik GmbH, coupled with an optical fibre to a lens-based telescope mounted on a sun tracker for measuring direct UV irradiance in the ultraviolet wavelength band between 305 to 345 nm. Two different algorithms are developed for retrieving TCO from these spectral measurements: (1) TCO retrieved by a least-squares-fit algorithm (LSF) and (2) a custom-double-ratio (CDR) technique using four specifically selected wavelengths from the spectral measurements. The double-ratio technique is analogous to the retrieval algorithm applied for the Dobson instruments and the Brewer instruments but is adopted here for TCO retrieval with Koherent. The instrument was calibrated in two different ways: (a) absolute calibration of the spectra using the portable reference for ultraviolet radiation QASUME for the LSF retrieval and (b) relative calibration of the extraterrestrial constant (ETC) of the CDR retrieval by minimising the slope between air mass and the relative differences of TCO from QASUME and Koherent. This adjustment of the ETC allows the instrument to be calibrated with standard TCO reference instruments during calibration campaigns, such as a double-monochromator Brewer. A 2-year comparison in Davos, Switzerland, between Koherent and the Brewer 156 (double monochromator) shows that TCO derived from Koherent and the Brewer 156 agree, on average, over the entire period within 0.7 % for all retrievals in terms of offset. The performance in terms of slant path depends on the selected retrieval and the applied corrections. The stray-light-corrected LSF retrieval exhibits a smaller slant path dependency than the CDR retrieval and performs almost as well as a double-monochromator system. The slant path dependency of the CDR is comparable to the slant path dependency of a single Brewer monochromator. The combination of both retrievals leads to performance with an offset close to zero compared to Brewer 156, a seasonal amplitude of the relative difference of 0.08 % and a slant path dependency of maximum 1.64 %, which is similar to other standard TCO instruments such as single Brewer or Dobson. Applying the double-ratio technique by selecting the wavelengths and slit functions from Brewer and Dobson, respectively, allows for the determination of the effective ozone temperature with an uncertainty of 3 K in terms of daily averages. With the improved TCO retrieval, Koherent serves as a new low-maintenance instrument which could also be used to monitor TCO at remote sites. The TCO retrieval presented here may be applied to other array-based spectroradiometers, providing direct spectral measurements in the ultraviolet wavelength band.

Funder

European Space Agency

Publisher

Copernicus GmbH

Subject

Atmospheric Science

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3