A cryogen-free, semi-automated apparatus for bullet-dynamic nuclear polarization with improved resolution
-
Published:2021-11-11
Issue:2
Volume:2
Page:815-825
-
ISSN:2699-0016
-
Container-title:Magnetic Resonance
-
language:en
-
Short-container-title:Magn. Reson.
Author:
Kouřil Karel, Gramberg Michel, Jurkutat Michael, Kouřilová Hana, Meier BennoORCID
Abstract
Abstract. In dissolution-dynamic nuclear polarization, a hyperpolarized solid is dissolved with a jet of hot solvent. The solution is then transferred to a secondary magnet, where spectra can be recorded with improved sensitivity.
In bullet-dynamic nuclear polarization this order is reversed. Pressurized gas is used to rapidly transfer the hyperpolarized solid to the secondary magnet, and the hyperpolarized solid is dissolved only upon arrival. A potential advantage of this approach is that it may avoid excessive dilution and the associated signal loss, in particular for small sample quantities. Previously, we have shown that liquid-state NMR spectra with polarization levels of up to 30 % may be recorded within less than 1 s after the departure of the hyperpolarized solid from the polarizing magnet. The resolution of the recorded spectra however was limited. The system consumed significant amounts of liquid helium, and substantial manual work was required in between experiments to prepare for the next shot. Here, we present a new bullet-DNP (dynamic nuclear polarization) system that addresses these limitations.
Funder
H2020 European Research Council Helmholtz-Gemeinschaft Engineering and Physical Sciences Research Council
Publisher
Copernicus GmbH
Reference49 articles.
1. Ardenkjær-Larsen, J. H.: On the Present and Future of Dissolution-DNP,
J. Magn. Reson., 264, 3–12, https://doi.org/10.1016/j.jmr.2016.01.015,
2016. a 2. Ardenkjær-Larsen, J. H.: Introduction to dissolution DNP: Overview,
instrumentation, and human applications, Handbook of High Field Dynamic
Nuclear Polarization, John Wiley & Sons, Chichester, UK, p. 239, 2019. a 3. Ardenkjær-Larsen, J. H., Fridlund, B., Gram, A., Hansson, G., Hansson, L.,
Lerche, M. H., Servin, R., Thaning, M., and Golman, K.: Increase in
Signal-To-Noise Ratio of > 10 000 Times in Liquid-State NMR,
P. Natl. Acad. Sci. USA, 100, 10158–10163,
https://doi.org/10.1073/pnas.1733835100, 2003. a 4. Ardenkjaer-Larsen, J.-H., Boebinger, G. S., Comment, A., Duckett, S., Edison,
A. S., Engelke, F., Griesinger, C., Griffin, R. G., Hilty, C., Maeda, H.,
Parigi, G., Prisner, T., Ravera, E., van Bentum, J., Vega, S., Webb, A.,
Luchinat, C., Schwalbe, H., and Frydman, L.: Facing and Overcoming
Sensitivity Challenges in Biomolecular NMR Spectroscopy, Angew. Chem.
Int. Edit., 54, 9162–9185, https://doi.org/10.1002/anie.201410653, 2015. a 5. Ardenkjær-Larsen, J. H., Bowen, S., Petersen, J. R., Rybalko, O., Vinding, M. S., Ullisch, M., and Nielsen, N. C.: Cryogen-Free Dissolution Dynamic Nuclear Polarization Polarizer Operating At 3.35 T, 6.70 T, and 10.1 T, Magn. Reson. Med., 81, 2184, https://doi.org/10.1002/mrm.27537, 2018. a
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|