Abstract. A continuous source for the generation of pure HONO mixtures was developed
and characterized, which is based on the Henry's law solubility of HONO in
acidic aqueous solutions. With the help of a peristaltic pump, diluted
nitrite and sulfuric acid solutions are mixed in a temperature-controlled
stripping coil, which is operated with pure nitrogen or synthetic air at gas
flow rates of 0.5–2 L min−1. Caused by the acidic conditions of the
aqueous phase (pH ≈ 2.5), nitrite is almost completely converted
into HONO, which partitions to the gas phase limited by its known solubility in water. The source shows a fast time response of ∼ 2 min (0 %–90 %) at higher concentrations and an excellent long-term stability (2σ noise < 1 %). The HONO emission of the source perfectly
correlates with the nitrite concentration from the sub-ppb range up to 500 ppb. The rate of NOx formation increases quadratically with the HONO concentration from non-detectable values at atmospheric relevant HONO
concentrations reaching a NOx content of 1.6 % at 500 ppb. A general equation based on Henry's law is developed, whereby the HONO concentration of the source can be calculated using measured experimental parameters, i.e.
nitrite concentration, liquid flow rates, gas flow rate, pH of the solution,
and temperature of the stripping coil. In the equation, the known Henry's
law constant of HONO in sulfuric acid solutions is used. For the calculation
of the effective Henry's law constant, the acid dissociation equilibrium of
HONO / nitrite is used as a variable to adjust the theoretical HONO
concentration to the measured values. From the average of all experimental
data the equilibrium of HONO / nitrite is described well by
pKa=1021.53/T-0.449. The pKa of 3.0 ± 0.1 (1σ) at 25 ∘C is in good agreement with the range of 2.8–3.28 published in former studies. A standard deviation between all measured and theoretical HONO concentrations of only ±3.8 % was observed, and a conservative
upper-limit accuracy of the HONO concentration of better 10 % is
estimated. Thus, for the first time, a stable HONO source is developed
which can be used for the absolute calibration of HONO instruments.