Drone-towed controlled-source electromagnetic (CSEM) system for near-surface geophysical prospecting: on instrument noise, temperature drift, transmission frequency, and survey set-up
-
Published:2022-12-08
Issue:2
Volume:11
Page:435-450
-
ISSN:2193-0864
-
Container-title:Geoscientific Instrumentation, Methods and Data Systems
-
language:en
-
Short-container-title:Geosci. Instrum. Method. Data Syst.
Author:
Vilhelmsen Tobias BjergORCID, Døssing Arne
Abstract
Abstract. Drone-borne controlled-source electromagnetic (CSEM) systems combine the mobility of airborne systems with the high subsurface resolution in ground
systems. As such, drone-borne systems are beneficial at sites with poor accessibility and in areas where high resolution is needed, e.g. for
archaeological or subsurface pollution investigations. However, drone-borne CSEM systems are associated with challenges, which are not observed to
the same degree in airborne or ground surveys. In this paper, we explore some of these challenges based on an example of a new drone-towed CSEM
system. The system deploys a multi-frequency broadband electromagnetic sensor (GEM-2 uncrewed aerial vehicle, UAV), which is towed 6 m below a drone in a towing-bird
configuration together with a NovAtel GNSS–IMU (global navigation satellite system–inertial measurement unit) unit, enabling centimetre-level position precision and orientation. The results of a number of
controlled tests of the system are presented together with data from an initial survey at Falster (Denmark), including temperature drift, altitude
vs. signal, survey mode signal dependency, and the effect of frequency choice on noise. The test results reveal the most critical issues for our
system and issues that are likely encountered in similar drone-towed CSEM set-ups. We find that small altitude variations (± 0.5 m)
along our flight paths drastically change the signal, and a local height vs. signal correlation is needed to correct near-surface drone-towed CSEM
data. The highest measured impact was −46.2 ppm cm−1 for a transmission frequency of 91 kHz. We also observe a significant increase in the
standard deviation of the noise level up to 500 % when going from one transmission frequency to five. We recommend not to use more than three
transmission frequencies, and the lowest transmission frequencies should be as high as the application allows it. Finally, we find a strong
temperature dependency (up to 32.2 ppm∘C-1), which is not accounted for in
the instrumentation.
Funder
A.P. Møller og Hustru Chastine Mc-Kinney Møllers Fond til almene Formaal
Publisher
Copernicus GmbH
Subject
Atmospheric Science,Geology,Oceanography
Reference31 articles.
1. Abdu, H., Robinson, D., and Jones, S. B.:
Comparing bulk soil electrical conductivity determination using the DUALEM-1S and EM38-DD electromagnetic induction instruments, Soil Sci. Soc. Am. J., 71, 189–196, 2007. a 2. Altdorff, D., Schliffke, N., Riedel, M., Schmidt, V., van der Kruk, J., Vereecken, H., Stoll, J., and Becken, M.:
UAV-borne electromagnetic induction and ground-penetrating radar measurements: a feasibility test, Water Resour. Res., 42, W11403, 2014. a 3. Bjella, K. L., Astley, B. N., and North, R. E.:
Geophysics for Military Construction Projects, Tech. rep., Engineer Research and Development Center, Vicksburg, MS, 2010. a 4. Bjerg, T., da Silva, E. L. S., and Døssing, A.:
Investigation of UAV noise reduction for electromagnetic induction surveying, in: NSG2020 3rd Conference on Geophysics for Mineral Exploration and Mining, 8 December 2020, vol. 2020, pp. 1–5, European Association of Geoscientists & Engineers, 2020. a 5. Callegary, J. B., Ferré, T. P., and Groom, R.: