System for <i>δ</i><sup>13</sup>C–CO<sub>2</sub> and <i>x</i>CO<sub>2</sub> analysis of discrete gas samples by cavity ring-down spectroscopy
-
Published:2017-11-22
Issue:11
Volume:10
Page:4507-4519
-
ISSN:1867-8548
-
Container-title:Atmospheric Measurement Techniques
-
language:en
-
Short-container-title:Atmos. Meas. Tech.
Author:
Dickinson DaneORCID, Bodé SamuelORCID, Boeckx Pascal
Abstract
Abstract. A method was devised for analysing small discrete gas samples (50 mL syringe) by cavity ring-down spectroscopy (CRDS). Measurements were accomplished by inletting 50 mL syringed samples into an isotopic-CO2 CRDS analyser (Picarro G2131-i) between baseline readings of a reference air standard, which produced sharp peaks in the CRDS data feed. A custom software script was developed to manage the measurement process and aggregate sample data in real time. The method was successfully tested with CO2 mole fractions (xCO2) ranging from < 0.1 to > 20 000 ppm and δ13C–CO2 values from −100 up to +30 000 ‰ in comparison to VPDB (Vienna Pee Dee Belemnite). Throughput was typically 10 samples h−1, with 13 h−1 possible under ideal conditions. The measurement failure rate in routine use was ca. 1 %. Calibration to correct for memory effects was performed with gravimetric gas standards ranging from 0.05 to 2109 ppm xCO2 and δ13C–CO2 levels varying from −27.3 to +21 740 ‰. Repeatability tests demonstrated that method precision for 50 mL samples was ca. 0.05 % in xCO2 and 0.15 ‰ in δ13C–CO2 for CO2 compositions from 300 to 2000 ppm with natural abundance 13C. Long-term method consistency was tested over a 9-month period, with results showing no systematic measurement drift over time. Standardised analysis of discrete gas samples expands the scope of application for isotopic-CO2 CRDS and enhances its potential for replacing conventional isotope ratio measurement techniques. Our method involves minimal set-up costs and can be readily implemented in Picarro G2131-i and G2201-i analysers or tailored for use with other CRDS instruments and trace gases.
Publisher
Copernicus GmbH
Subject
Atmospheric Science
Reference52 articles.
1. Albanito, F., McAllister, J. L., Cescatti, A., Smith, P., and Robinson, D.: Dual-chamber measurements of δ13C of soil-respired CO2 partitioned using a field-based three end-member model, Soil Biol. Biochem., 47, 106–115, 2012. 2. Bai, M., Köstler, M., Kunstmann, J., Wilske, B., Gattinger, A., Frede, H.-G., and Breuer, L.: Biodegradability screening of soil amendments through coupling of wavelength-scanned cavity ring-down spectroscopy to multiple dynamic chambers, Rapid Commun. Mass Sp., 25, 3683–3689, 2011. 3. Berryman, E. M., Marshall, J. D., Rahn, T., Cook, S. P., and Litvak, M.: Adaptation of continuous-flow cavity ring-down spectroscopy for batch analysis of δ13C of CO2 and comparison with isotope ratio mass spectrometry, Rapid Commun. Mass Sp., 25, 2355–2360, 2011. 4. Brewer, P. J., Brown, R. J., Miller, M. N., Miñarro, M. D., Murugan, A., Milton, M. J., and Rhoderick, G. C.: Preparation and Validation of Fully Synthetic Standard Gas Mixtures with Atmospheric Isotopic Composition for Global CO2 and CH4 Monitoring, Anal. Chem., 86, 1887–1893, 2014. 5. Christiansen, J. R., Outhwaite, J., and Smukler, S. M.: Comparison of CO2, CH4 and N2O soil-atmosphere exchange measured in static chambers with cavity ring-down spectroscopy and gas chromatography, Agr. Forest. Meteorol., 211–212, 48–57, 2015.
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|