Continuous in-line monitoring of electrolyte concentrations in extracorporeal circuits for individualization of dialysis treatment
-
Published:2018-10-29
Issue:2
Volume:7
Page:559-567
-
ISSN:2194-878X
-
Container-title:Journal of Sensors and Sensor Systems
-
language:en
-
Short-container-title:J. Sens. Sens. Syst.
Author:
Berger MarcORCID, Faulstich Michael, Perl Thorsten, Zimmermann Stefan
Abstract
Abstract. One objective of dialysis treatment is to normalize the blood
plasma electrolytes and remove waste products such as urea and creatinine
from blood. However, due to a shift in plasma osmolarity, a rapid or
excessive change of the electrolytes can lead to complications like
cardiovascular instability, overhydrating of cells, disequilibrium syndrome
and cardiac arrhythmias. Especially for critical ill patients in intensive
care unit with sepsis or multi-organ failure, any additional stress has to be
avoided. Since the exchange velocity of the electrolytes mainly depends on
the concentration gradients across the dialysis membrane between blood and
dialysate, it can be controlled by an individualized composition of dialysate
concentrations. In order to obtain a precise concentration gradient with the
individualized dialysate, it is necessary to continuously monitor the plasma
concentrations. However, with in-line sensors, the required hemocompatibility
is often difficult to achieve. In this work, we present a concept for
continuous in-line monitoring of electrolyte concentrations using
ion-selective electrodes separated from the blood flow by a dialysis
membrane, and therefore meeting the fluidic requirements for
hemocompatibility. First investigations of hemocompatibility with
reconfigured human blood show no increased hemolysis caused by the measuring
system. With this concept, it is possible to continuously measure the plasma
concentrations with a relative error of less than 0.5 %.
Publisher
Copernicus GmbH
Subject
Electrical and Electronic Engineering,Instrumentation
Reference44 articles.
1. Bagshaw, S. M., George, C., and Bellomo, R.: Early acute kidney injury and sepsis: A multicentre evaluation, Crit. Care, 12, R47, https://doi.org/10.1186/cc6863, 2008. 2. Baue, A. E., Faist, E., and Fry, D. E.: Multiple Organ Failure: Pathophysiology, Prevention, and Therapy, N, Springer New York, New York, 1, online resource 722, 2000. 3. Bellomo, R., Kellum, J. A., and Ronco, C.: Acute kidney injury, The Lancet, 380, 756–766, https://doi.org/10.1016/S0140-6736(11)61454-2, 2012. 4. Buemi, M., Aloisi, E., Coppolino, G., Loddo, S., Crascì, E., Aloisi, C., Barillà, A., Cosentini, V., Nostro, L., Caccamo, C., Floccari, F., Romeo, A., Frisina, N., and Teti, D.: The effect of two different protocols of potassium haemodiafiltration on QT dispersion, Nephrol. Dial. Transpl., 20, 1148–1154, https://doi.org/10.1093/ndt/gfh770, 2005. 5. Bundesärztekammer: Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen, Deutsches Ärzteblatt, 111, A 1583–A 1618, 2014.