The sensitivity of benzene cluster cation chemical ionization mass spectrometry to select biogenic terpenes
-
Published:2018-06-07
Issue:6
Volume:11
Page:3251-3262
-
ISSN:1867-8548
-
Container-title:Atmospheric Measurement Techniques
-
language:en
-
Short-container-title:Atmos. Meas. Tech.
Author:
Lavi Avi, Vermeuel Michael P., Novak Gordon A., Bertram Timothy H.ORCID
Abstract
Abstract. Benzene cluster cations are a sensitive and selective reagent ion for
chemical ionization of select biogenic volatile organic compounds. We have
previously reported the sensitivity of a field deployable chemical ionization
time-of-flight mass spectrometer (CI-ToFMS), using benzene cluster cation ion
chemistry, for detection of dimethyl sulfide, isoprene and α-pinene.
Here, we present laboratory measurements of the sensitivity of the same
instrument to a series of terpenes, including isoprene, α-pinene,
β-pinene, D-limonene, ocimene, β-myrcene, farnesene, α-humulene, β-caryophyllene, and isolongifolene at atmospherically
relevant mixing ratios (< 100 pptv). In addition, we determine the
dependence of CI-ToFMS sensitivity on the reagent ion neutral delivery
concentration and water vapor concentration. We show that isoprene is
primarily detected as an adduct (C5H8 ⋅ C6H6+) with a sensitivity ranging between 4 and 10 ncps ppt−1,
which depends strongly on the reagent ion precursor concentration,
de-clustering voltages, and specific humidity (SH). Monoterpenes are detected
primarily as the molecular ion (C10H16+) with an average
sensitivity, across the five measured compounds, of 14 ± 3 ncps ppt−1 for SH between 7 and 14 g kg−1, typical of the boreal
forest during summer. Sesquiterpenes are detected primarily as the molecular
ion (C15H24+) with an average sensitivity, across the four
measured compounds, of 9.6 ± 2.3 ncps ppt−1, that is also
independent of specific humidity. Comparable sensitivities across broad
classes of terpenes (e.g., monoterpenes and sesquiterpenes), coupled to the
limited dependence on specific humidity, suggest that benzene cluster cation
CI-ToFMS is suitable for field studies of biosphere–atmosphere interactions.
Funder
Division of Atmospheric and Geospace Sciences Office of Science Camille and Henry Dreyfus Foundation
Publisher
Copernicus GmbH
Subject
Atmospheric Science
Reference44 articles.
1. Allan, J. D., Alfarra, M. R., Bower, K. N., Coe, H., Jayne, J. T., Worsnop, D. R., Aalto, P. P., Kulmala, M., Hyötyläinen, T.,
Cavalli, F., and Laaksonen, A.: Size and composition measurements of background aerosol and new particle growth in a Finnish forest
during QUEST 2 using an Aerodyne Aerosol Mass Spectrometer, Atmos. Chem. Phys., 6, 315–327, https://doi.org/10.5194/acp-6-315-2006, 2006. 2. Bertram, T. H., Kimmel, J. R., Crisp, T. A., Ryder, O. S., Yatavelli, R. L. N., Thornton, J. A., Cubison, M. J., Gonin, M.,
and Worsnop, D. R.: A field-deployable, chemical ionization time-of-flight mass spectrometer, Atmos. Meas. Tech., 4,
1471–1479, https://doi.org/10.5194/amt-4-1471-2011, 2011. 3. Bieri, G., Asbrink, L., and Vonniessen, W.: 30.4-Nm He(Ii)
Photoelectron-Spectra of Organic-Molecules 4. Fluoro-Compounds (C,H,F),
J. Electron Spectrosc., 23, 281–322, 1981. 4. Chan, W. F., Cooper, G., and Brion, C. E.: The Electronic-Spectrum of Water
in the Discrete and Continuum Regions – Absolute Optical
Oscillator-Strengths for Photoabsorption (6–200 Ev), Chem. Phys., 178,
387–400, 1993. 5. Chipot, C., Jaffe, R., Maigret, B., Pearlman, D. A., and Kollman, P. A.:
Benzene dimer: A good model for pi-pi interactions in proteins? A comparison
between the benzene and the toluene dimers in the cas phase and in an
aqueous solution, J. Am. Chem. Soc., 118,
11217–11224, 1996.
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|