Measurement of the absolute spectral responsivity in the mid-infrared based on the cryogenic electrical substitution radiometer and an optimized thermopile detector
-
Published:2019-05-15
Issue:1
Volume:8
Page:195-205
-
ISSN:2194-878X
-
Container-title:Journal of Sensors and Sensor Systems
-
language:en
-
Short-container-title:J. Sens. Sens. Syst.
Author:
Pohl Tobias,Meindl Peter,Johannsen Uwe,Taubert Dieter,Werner Lutz
Abstract
Abstract. The Physikalisch-Technische Bundesanstalt (PTB) expanded its capabilities of
the absolute measurement of radiant power to the spectral range of the
mid-infrared (MIR) by implementing additional MIR laser radiation sources at
one of the PTB's cryogenic electrical substitution radiometer facilities.
This extension enables absolute calibrations of the spectral responsivity of
detectors in the MIR traceable to the International System of Units (SI). The thermopile detector TS-76 was characterized and calibrated in view of its
spectral responsivity s(λ) in the wavelength range between 1.5 and
10.6 µm at the expanded cryogenic electrical substitution
radiometer facility. The relative standard measurement uncertainty was
significantly reduced to 1.4 % by developing an optimized and thermally
stabilized detector housing
design. The TS-76 was established as a mid-infrared transfer detector for the
SI traceable measurement of radiant power and the dissemination of the
spectral responsivity s(λ) in the MIR.
Publisher
Copernicus GmbH
Subject
Electrical and Electronic Engineering,Instrumentation
Reference21 articles.
1. Adibekyan, A., Kononogova, E., Monte, C., and Hollandt, J.: High-Accuracy
Emissivity Data on the Coatings Nextel 811-21, Herberts 1534, Aeroglaze Z306
and Acktar Fractal Black, Int. J. Thermophys., 38, 1–14, 2017. 2. Allison, R. S., Johnston, J. M., Craig, G., and Jennings, S.: Airborne
Optical and Thermal Remote Sensing for Wildfire Detection and Monitoring,
Sensors-Basel, 16, 1310, https://doi.org/10.3390/s16081310, 2016. 3. BIPM (Bureau International des Poids et Mesures): Guide to the expression of
uncertainty in measurement (GUM), International Organization for
Standardization, Sèvres, 1995. 4. Budzier, H. and Gerlach, G.: Thermische Infrarotsensoren – Grundlagen
für Anwender, Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim, 2010. 5. Gottwald, A., Kroth, U., Krumrey, M., Richter, M., Scholze, F., and Ulm, G.:
The PTB high-accuracy spectral responsivity scale in the VUV and x-ray range,
Metrologia, 43, S125–S129, 2006.