Development of a gas chromatography system coupled to a metal-oxide semiconductor (MOS) sensor, with compensation of the temperature effects on the column for the measurement of ethene
-
Published:2023-08-08
Issue:2
Volume:12
Page:215-223
-
ISSN:2194-878X
-
Container-title:Journal of Sensors and Sensor Systems
-
language:en
-
Short-container-title:J. Sens. Sens. Syst.
Author:
Koehne Maximilian, Schmidt Christopher, Singh Satnam, Grasskamp AndreasORCID, Sauerwald Tilman, Zeh Gina
Abstract
Abstract. A possible way to reduce the size and complexity of
common gas chromatography (GC) systems is the economization of the column
temperature regulation system. To this end, a temperature compensation
method was developed and validated on a benchtop GC-PDD (pulsed discharge
detector) with ethene. An in-house-developed algorithm correlates the
retention index of a test gas to the retention index of a previously
selected reference gas. To investigate further methods of cost reduction,
commercial gas sensors were tested as cheap, sensitive, and versatile
detectors. Therefore, CO2 was chosen as a naturally occurring reference gas, while ethene was chosen as a maturity marker for climacteric fruits and
hence as a test gas. A demonstrator, consisting of a simple
syringe injection system, a PLOT (porous layer open tubular) column boxed in
a polystyrene-foam housing, a commercial MOS (metal-oxide semiconductor)
sensor for the test gas, and a CO2-specific IR (infrared) sensor, was
used to set up a simple GC system and to apply this method on
test measurements. Sorption parameters for ethene and CO2 were
determined via a van 't Hoff plot, where the entropy S was −11.982 J mol−1 K−1 ΔSEthene0 and 1.351 J mol−1 K−1 ΔSCarbondioxide0, and the enthalpy H was −20.622 kJ mol−1 ΔHEthene0 and −14.792 kJ mol−1 ΔHCarbondioxide0, respectively. Ethene (100 ppm) measurements
revealed a system-specific correction term of 0.652 min. Further
measurements of ethene and interfering gases revealed a mean retention time
for ethene of 3.093 min; the mean predicted retention time is 3.099 min. The
demonstrator was able to identify the test gas, ethene, as a function of
the reference gas, CO2, in a first approach, without a column heating
system and in a gas mixture by applying a temperature compensation algorithm
and a system-specific holdup time correction term.
Funder
Bayerische Staatsministerium für Wirtschaft, Landesentwicklung und Energie
Publisher
Copernicus GmbH
Subject
Electrical and Electronic Engineering,Instrumentation
Reference40 articles.
1. Agilent Technologies: Agilent J&W Column Performance Summary GS-Q, Light Hydrocarbon,
https://www.agilent.com/en/products/gas-chromatography/gc-columns/capillary/gs-q/gp43071, last access 31 July 2023. 2. Azzouz, I. and Bachari, K.: MEMS Devices for Miniaturized Gas
Chromatography, IntechOpen, 149–169,
https://doi.org/10.5772/intechopen.74020, 2018. 3. Baur, T., Schultealbert, C., Schütze, A., and Sauerwald, T.: Novel method for the detection of short trace gas pulses with metal oxide semiconductor gas sensors, J. Sens. Sens. Syst., 7, 411–419, https://doi.org/10.5194/jsss-7-411-2018, 2018a. 4. Baur, T., Schultealbert, C., Schütze, A., and Sauerwald, T.: Device for the
detection of short trace gas pulses, tm – Technisches Messen, 496–503,
https://doi.org/10.1515/teme-2017-0137, 2018b. 5. Cagliero, C., Galli, S., Galli, M., Elmi, I., Belluce, M., Zampolli, S., Sgorbini,
B., Rubiolo, P., and Bicchi, C.: Conventional and enantioselective gas
chromatography with microfabricated planar columns for analysis of
real-world samples of plant volatile fraction, J. Chromatogr. A,
1429, 329–339, https://doi.org/10.1016/j.chroma.2015.12.037, 2016.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|