Abstract. We present the development of a chemical ionization mass spectrometer ion source specifically designed for in situ measurements of trace gases in
the upper troposphere and lower stratosphere. The ion source utilizes a
commercially available photoionization krypton lamp, primarily emitting
photons in the vacuum ultraviolet (VUV) region at wavelengths of 124 and 117 nm (corresponding to energies of 10 and 10.6 eV, respectively), coupled to a
commercially available Vocus proton transfer reaction mass spectrometer. The
VUV ion source can produce both negative and positive reagent ions; however,
here we primarily focus on generating iodide anions (I−). The
instrument's drift tube (also known as ion–molecule reactor) operates at
pressures between 2 and 10 mbar, which facilitates ambient sampling at
atmospheric pressures as low as 50 mbar. The low operating pressure reduces
secondary ion chemistry that can occur in iodide chemical ionization mass spectrometry (CIMS). It also allows the
addition of water vapor to the drift tube to exceed typical ambient humidity
by more than 1 order of magnitude, significantly reducing ambient humidity
dependence of sensitivities. An additional benefit of this ion source and
drift tube is a 10- to 100-fold reduction in nitrogen consumed during
operation relative to standard I− ion sources, resulting in
significantly reduced instrument weight and operational costs. In iodide
mode, sensitivities of 76 cps ppt−1 for nitric acid, 35 cps ppt−1 for Br2
and 8.9 cps ppt−1 for Cl2 were achieved. Lastly, we demonstrate that this
ion source can generate benzene (C6H6+) and ammonium
(NH4+) reagent ions to expand the number of detected atmospheric
trace gases.
Funder
National Oceanic and Atmospheric Administration
Publisher
Copernicus GmbH
Subject
Atmospheric Science
Reference38 articles.
1. Bertram, T. H., Kimmel, J. R., Crisp, T. A., Ryder, O. S., Yatavelli, R. L. N., Thornton, J. A., Cubison, M. J., Gonin, M., and Worsnop, D. R.: A field-deployable, chemical ionization time-of-flight mass spectrometer, Atmos. Meas. Tech., 4, 1471–1479, https://doi.org/10.5194/amt-4-1471-2011, 2011. 2. Breitenlechner, M., Novak, G. A., Neuman, J. A., Rollins, A. W., and Veres, P. R.: Index of /groups/csl7/datasets/data/ivocus, NOAA [data set], https://csl.noaa.gov/groups/csl7/datasets/data/ivocus/, last acces: 3 March 2022. 3. Caldwell, G. W., Masucci, J. A., and Ikonomou, M. G.: Negative ion chemical
ionization mass spectrometry – binding of molecules to bromide and iodide
anions, J. Org. Mass Spectrom, 24, 8–14,
https://doi.org/10.1002/oms.1210240103, 1989. 4. Crounse, J. D., McKinney, K. A., Kwan, A. J., and Wennberg, P. O.:
Measurement of Gas-Phase Hydroperoxides by Chemical Ionization Mass
Spectrometry, Anal. Chem., 78, 6726–6732,
https://doi.org/10.1021/ac0604235, 2006. 5. de Gouw, J. and Warneke, C.: Measurements of Volatile Organic Compounds in
the Earth's Atmosphere using Proton-Transfer-Reaction Mass Spectrometry,
Mass Spectrom. Rev., 26, 223–257, https://doi.org/10.1002/mas.20119, 2007.
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|