Improving Decreased Heater–Cooler Efficiency as a Result of Heater–Cooler Infection Control Strategy

Author:

Blakey Adam K.,Holt David W.

Abstract

Heater–cooler units (HCUs) play a vital role in temperature management during cardiopulmonary bypass. In recent years, HCUs have been shown to play a significant role in the propagation of bacteria causing patient infection and significant harm. As a result, various institutions across the world have begun moving the HCU either far away or outside of the operative theater entirely. The purpose of this study was to examine the effect that the increased length of HCU water lines have on the ability of the device to heat and cool. We hypothesized that the increase in water line distance leads to a decrease in HCU efficiency and that insulating the water lines would blunt the effect of this increase in distance. Five water line conditions were compared under two cooling and two warming ranges. Short water lines, long water lines, and long water lines with foam, rubber, or tape insulation were compared. Cooling from an arterial line temperature of 26.7–19.7°C showed no difference between conditions with the exception that every long line condition takes significantly longer to cool than short water lines. Cooling from 35.6 to 28.6°C revealed that all insulations reduce the cooling time compared with long water lines without insulation, but only foam insulation reduces to the level of the short water lines. During warming conditions, all insulations reduced the warming time compared with long uninsulated water lines, but none were comparable with short water lines. Increased water line length leads to a decrease in HCU efficiency. Insulation is effective at increasing efficiency of long water lines, but only at warmer temperatures and not to the level of short water lines. Only foam-insulated long water lines were able to match the efficiency of short water lines, but only across a single temperature range.

Publisher

EDP Sciences

Subject

Cardiology and Cardiovascular Medicine,Health Professions (miscellaneous),Medicine (miscellaneous)

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3