Transition from Hemochron Response to Hemochron Signature Elite Activated Clotting Time Devices in a Congenital Cardiac Surgery Practice

Author:

Matte Gregory S.,Howe Robert J.,Ibla Juan,Emani Sirisha,Emani Sitaram M.

Abstract

Heparin is the primary anticoagulant used during cardiac surgery to prevent thrombosis due to cardiopulmonary bypass (CPB)–related activation of the hemostatic system. The efficacy of heparin in the operating room is generally determined by activated clotting time (ACT) point-of-care tests performed throughout the procedure. In an effort to transition to the Hemochron Elite which requires approximately 1/10th the sampling volume of blood, we conducted a prospective study in 260 pediatric patients undergoing CPB. ACT tests were performed during CPB with a total of 260 pre-bypass and 1,117 on-bypass ACT values recorded. All samples were run simultaneously on both ACT devices. Several therapeutic cut-off possibilities ranging from >380 to >480 seconds were evaluated to ascertain the ACT level on the Elite device which best correlated with results from the Response device. Linear regression was used to determine correlation. The correlation between the two methods was moderate with a Pearsonrof .6 and .4 for pre-bypass bolus ACT values and on-bypass ACT values, respectively. As the therapeutic ACT cut-off values were lowered from 480 to 380 seconds on the Elite device relative to the Response device (>480 seconds) for the on-bypass heparin samples, more patients would be under-dosed (incidence rising from 1 to 2%) and fewer patients would be overdosed (incidence decreasing from 32 to 5%) and the percent correlation between devices increased from 67 to 93%. A similar trend was observed with the pre-bypass heparin bolus samples. There was no significant effect of temperature on the ACT values comparing both devices. A therapeutic ACT value of >400 seconds for CPB with the Hemochron Elite device reasonably approximates a therapeutic ACT value of >480 seconds on the Hemochron Response device in our congenital cardiac surgery practice. Transitioning to the Elite device significantly reduces the overall sampling volume required for ACT monitoring during cardiac surgery.

Publisher

EDP Sciences

Subject

Cardiology and Cardiovascular Medicine,Health Professions (miscellaneous),Medicine (miscellaneous)

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. How We Would Treat Our Own Hypoplastic Left Heart Syndrome Neonate for Stage 1 Surgery;Journal of Cardiothoracic and Vascular Anesthesia;2023-04

2. 体外循環中ヘパリン濃度測定による抗凝固管理下で開心術を施行した抗リン脂質抗体症候群合併の1例;Japanese Journal of Cardiovascular Surgery;2023-01-15

3. Publication Ethics;The Journal of ExtraCorporeal Technology;2019-12

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3