Author:
Broman Lars Mikael,Eriksson Christer,Frenckner Björn,Palmér Kenneth
Abstract
In extracorporeal membrane oxygenation (ECMO), blood is drained from the patient, and pumped through a membrane oxygenator/lung (ML) for gas exchange and then back to the patient. For monitoring blood gases, samples may be sampled downstream from the ML. This exposes the patient for embolization risk (air/clot) when the stopcocks are flushed. For safe sampling procedures, the Loop was introduced. It is a constant low-flow arteriovenous shunt (AVS) used preferably in venoarterial ECMO. It is composed of three different length and diameter three-way stopcocks connected to the circuit just downstream the ML with its return upstream the pump. It offers safe arterial blood sampling and a simultaneous access point to the venous side of the circuit. Since its introduction, no patient complications have been reported to be accounted for by the Loop. The Loop is an AVS permitting a safe access point for post membrane blood sampling and for injections in the venous pre-pump limb. It has a low cost and is easy to install and maintain. It may be used in any ECMO configuration.
Subject
Cardiology and Cardiovascular Medicine,Health Professions (miscellaneous),Medicine (miscellaneous)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Learning about Journal Metrics;The Journal of ExtraCorporeal Technology;2019-09