Abstract
Objectives: Due to ongoing concern about femur anatomy-implant mismatches, this cross-sectional study aimed to create a geometric femur profile and used it to identify and quantify possible mismatches between femur anatomy and cephalomedullary nail dimensions. The work further aimed to assess whether patient demographics affect anatomy-implant coherence. Methods: One hundred skeletally mature complete femur computer tomography (CT) scans were collected and exported to software enabling landmark placement and measures with multiplanar reconstruction techniques. Results: Clinically relevant anatomy-implant discrepancies included the femur neck and shaft axis offset 6.1 ± 1.7 mm (95% CI [5.7–6.4]), femur radius of curvature 1.2 ± 0.3 m (95% CI [1.1–1.2]), femur anteversion 18.8 ± 9.2 (95% CI [16.9–20.6]). The implants reviewed in this study did not compensate for the femur neck and shaft axis offset and had a larger radius of curvature than the studied population. Clinically significant demographic geometry differences were not identified. Conclusion: There were discrepancies between femur anatomy and cephalomedullary nail implant design; however, no clinically significant femur feature inconsistency was identified among the demographic subgroups. Due to the identified anatomy-implant discrepancies, including the femur neck and shaft axis offset, we suggest that these measurements be considered for future implant design and surgical technique.