Author:
Dellepiane Gaia,Casolaro Pierluigi,Häffner Philipp Daniel,Mateu Isidre,Scampoli Paola,Voeten Naomi,Zyaee Elnaz,Braccini Saverio
Abstract
The production of theranostic radionuclides using solid targets is challenging and requires an accurate knowledge of the production crosssections as well as the energy, positioning and focusing of the beam. A research program is ongoing at the 18 MeV Bern medical cyclotron, equipped with a Solid Target Station (STS) and a 6 m Beam Transfer Line (BTL) ending in a separate bunker with independent access. A novel target coin was designed and built to irradiate compressed powder pellets, together with a compact focalization system to optimize the irradiation procedure. Furthermore, methods were developed to measure the beam energy, the production cross-sections and the EoB-activity.
Reference15 articles.
1. Braccini S. et al., Proceedings of Cyclotrons2019, Cape Town, South Africa, 127-131 (2019)
2. The new bern PET cyclotron, its research beam line, and the development of an innovative beam monitor detector
3. An Active Irradiation System with Automatic Beam Positioning and Focusing for a Medical Cyclotron
4. Dellepiane G. et al., Proceedings of the International Conference on Technology and Instrumentation in Particle Physics TIPP2021, Vancouver, Canada, on-line format, (2021)