Author:
Jakubowicz Michał,Rucki Mirosław,Siemiątkowski Zbigniew
Abstract
The paper describes the test rig dedicated for air gauge dynamical characteristics assessment. The computerised system enables measurement of the amplitudes of back-pressure pk dependent on the input signal circular frequency ω. Dedicated software performs full control on the calibration procedure, which consists of setting a rotational speed and registration of measuring signal, and further data processing. Circular frequency ω was gradually changed with the appropriate step, in order to obtain a series of frequencies in the range from 0.1 to 20 Hz. The response of a measurement system was registered as a sinusoidal curve which after smoothening and interpolation procedures provided an amplitude-frequency graph with its main characteristics, such as time constant and the frequency f0.95 that generated dynamic error 5%. It was demonstrated that sine input dynamic calibration corresponds with real conditions of the in-process measurement with air gauges.