Author:
Gouda M.M.,Abbas M.I.,Ismail A.O.,El-Khatib A.M.
Abstract
This work introduces a mathematical formula to calculate the attenuation coefficient of a cylindrical sample placed on the top of a well-shaped gamma-ray detector. The formula depends on calculations of the average path length of the photon through a sample and the effective solid angle. The obtained results using the introduced formula are compared with the NIST XCOM database and compared with the experimental work using two different-sized detectors and sources located in two different positions. A remarkable agreement between both was observed, which proves the validity of this formula.
Subject
Health, Toxicology and Mutagenesis,Public Health, Environmental and Occupational Health,Waste Management and Disposal,Safety, Risk, Reliability and Quality,Nuclear Energy and Engineering,Renewable Energy, Sustainability and the Environment
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献