Abstract
This study aims to establish radiation dose typical values (TV) for ventricular septal defect (VSD) percutaneous closures based on patient-, device- and procedure-related characteristics. Clinical and exposure data were retrospectively reviewed for 92 consecutive pediatric patients with VSD in whom transcatheter closure attempt was undertaken between March 2016 and December 2019 at a single reference center. TV were derived as the median value of total air kerma-area product (PKA,T), air kerma at the patient entrance reference point, fluoroscopy time and number of frames. Statistical analyses were conducted to investigate the exposure variability according to patient, device- and procedure-related characteristics. For (5–< 15 kg) patients, TV were 27 Gy · cm2 in term of PKA,T for venous vs. 9 Gy · cm2 for arterial approach. For (15–< 30 kg) patients, TV were 53 Gy · cm2 for muscular vs. 18 Gy · cm2 for perimembranous defects and 36 Gy · cm2 for AmplatzerTM vs. 14 Gy · cm2 for the LifetechTM device. Procedures conducted with the assistance of a fellow, venous approach, muscular VSD, AmplatzerTM devices, and procedural incidents increased the exposure levels significantly. The established TV provide benchmarks on patient exposure per procedure complexity and should help promote better patient radiation protection.
Subject
Health, Toxicology and Mutagenesis,Public Health, Environmental and Occupational Health,Waste Management and Disposal,Safety, Risk, Reliability and Quality,Nuclear Energy and Engineering,Renewable Energy, Sustainability and the Environment
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献