Affiliation:
1. University of Seville
2. University of Girona
3. Universidade de Lisboa
Abstract
This article presents an experimental investigation of biocompatible Polyether ether ketone (PEEK) polymeric sheet deformed using Single Point Incremental Forming (SPIF) at room temperature, with the objective of manufacturing a cranial medical implant. The investigation was performed in terms of formability and failure within the principal strains space, being the material Forming Limit Diagram (FLD) assessed by means of Nakajima tests. This material characterization allowed to obtain the formability limits at necking and at fracture of PEEK polymeric material. In addition, an experimental work plan in SPIF was performed in terms of spifability (or formability in SPIF) with the aim of determining the effect of the main process parameters on the spifability, modes of failure, and temperature, among others. As a result, an optimum set of parameters along with the established methodology were used for manufacturing a cranial medical implant made of this high-performance and biocompatible polymeric material. The work primarily shows the feasibility of manufacturing PEEK medical prostheses and implants using SPIF.
Publisher
Trans Tech Publications, Ltd.
Subject
Mechanical Engineering,Mechanics of Materials,General Materials Science
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献