Bioceramics are Not Bioinert: The Role of Oxide and Non-Oxide Bioceramics on the Oxidation of UHMWPE Components in Artificial Joints

Author:

Rondinella Alfredo1,Marin Elia1,McEntire Brian J.2,Bock Ryan2,Bal B. Sonny2,Zhu Wen Liang1,Yamamoto Kengo3,Pezzotti Giuseppe3

Affiliation:

1. Kyoto Institute of Technology

2. Amedica Corporation

3. Osaka University

Abstract

The following research is aimed at understanding the influence of Zirconia-Toughened Alumina (ZTA) and Silicon Nitride (Si3N4) on Ultra-High Molecular Weight Polyethylene (UHMWPE) acetabular liners. Bioceramic femoral heads were systematically tested against UHMWPE in controlled environment according to static/load-free coupling in hydrothermal environment, pin-on-ball wear testing, and hip-simulator wear testing. In addition, a retrieved ZTA femoral head has been analyzed and results have been compared to the simulations. Experimental results from X-ray photoelectron (XPS), cathodoluminescence (CL), Raman and Fourier-Transformed Infrared spectroscopy suggest that, despite conventional notions imply that bioceramics are inert, the surface chemistry of bioceramics was relevant to the oxidation rate of polyethylene liners. Non-biointertness could either be advantageous or disadvantageous toward polyethylene oxidation. The main reason resides in the peculiar chemical interactions between polyethylene and different ceramics, and, more specifically, depends on the direction of oxygen flow at the interface between the ceramic and the polymer. ZTA femoral heads were found to release a non-negligible amount of oxygen moieties from their surfaces, thus accelerating oxidative degradation of polyethylene. Conversely, Si3N4 ceramics exerted a protective role towards the polyethylene liner by scavenging oxygen from the tribolayer. The results of this work provide new insights into the interaction between bioceramics and polymers, which should also be considered when designing the next generation artificial hip joints with significantly elongated lifetimes.

Publisher

Trans Tech Publications, Ltd.

Subject

Mechanical Engineering,Mechanics of Materials,General Materials Science

Reference23 articles.

1. S.R. Knight, R. Aujla, S.P. Biswas, Total Hip Arthroplasty-over 100 years of operative history, Orthopedic reviews 3.2 (2011).

2. P.S. Christel, Biocompatibility of surgical-grade dense polycrystalline alumina, Clinical orthopaedics and related research, 282 (1992) 10-18.

3. Australian Orthopaedic Association National Joint Replacement Registry, Annual Report 2015. Available at: https://aoanjrr.sahmri.com/documents/10180/217745/Hip%20and%20Knee% 20Arthroplasty.

4. Swedish Hip Arthroplasty Register, Annual Report 2013. Available at: https://registercentrum.blob.core.windows.net/shpr/r/Annual-report-2013-HJnnK8Tie.

5. The New Zealand Joint Registry, Fifteen Year Report: January 1999 to December 2013. Available at: https://nzoa.org.nz/system/files/NZJR2014Report.pdf.