Abstract
Using Fourier transform infrared spectroscopy (FTIR), attenuated total reflection (ATR) technology and partial least square (PLS) method, the rapid quantification method of hemoglobin (HGB) in human soluble blood samples was established. Based on the distribution of samples’ HGB chemical value and absorbance on 1543 cm-1 which had the highest signal to noise ratio for HGB, all samples were divided into calibration set and prediction set for 50 times. PLS models were established for all divisions, based on the average data RMSEPAve, the stable optimal model was selected, the corresponding PLS factor, RMSEPAve and RP,Ave were 2, 6.81 g/L and 0.943 respectively.