Affiliation:
1. University of Oxford
2. University of Southampton
3. University of Limerick, Materials and Surface Science Institute
Abstract
Self-expanding stents are small medical devices used to treat vascular disease and are typically fabricated from a super-elastic, shape memory alloy known as Nitinol and have a fine mesh structure. This paper describes preliminary work on the application of Thermoelastic Stress Analysis (TSA) to Nitinol stents. Uniaxial tensile tests were conducted on thin tubes of Nitinol to characterise the material mechanical properties. TSA calibration exercises were conducted, which showed that Nitinol exhibits a non-uniform thermoelastic response through its elastic region that corresponded to the superelastic behaviour. Initial TSA demonstrated that a viable thermoelastic signal could be obtained from the stents. In high resolution tests the effect of motion and noise were considerable but it was still possible to obtain a readable thermoelastic signal.