In the late 1970s, based on good clinical results in total hip prostheses, the use of a
combination of alumina ceramics and ultra-high molecular weight polyethylene (UHMWPE) was
begun for total knee prostheses (TKPs) in order to reduce UHMWPE wear. In this study, to
examine the in vivo efficacy of the alumina ceramic bearing surfaces of TKPs, we compared
retrieved alumina ceramic TKPs with cobalt-chrome (Co-Cr) alloy TKPs by surface observations
and linear wear measurements. In scanning electron microscopic observations, many scratches due
to clinical use were observed only on the retrieved Co-Cr alloy femoral components. The damage
in the form of scratches on the articulating surface was linear and was produced by rubbing against
the Co-Cr alloy surface. The linear wear rate of the retrieved Co-Cr alloy TKPs was 0.027–0.358
mm/year. In contrast, the wear of the retrieved alumina ceramic TKPs was stably low and linear;
the linear wear rate was estimated to be 0.026 mm/year. The lower wear rate and milder nature of
wear observed in the TKPs with the combination of UHMWPE inserts and alumina ceramic femoral
components suggest the possibility of the retention of high performance even during prolonged
clinical use.
Publisher
Trans Tech Publications, Ltd.
Subject
Mechanical Engineering,Mechanics of Materials,General Materials Science
Reference4 articles.
1. H. Oonishi, T. Hasegawa (1981) Orthopedic Ceramic Implants 1: 157-160.
2. H. Oonishi, E. Tsuji, T. Mizukoshi, et al. (1991) Bioceramics 3: 137-145.
3. H. Oonishi, S.C. Kim, M. Kyomoto, et al. (2005) J Biomed Mater Res Appl Biomater 74: 754-759.
4. T. Nakanishi, K. Shikata, Y. Wang, et al. (2006) Key Eng Mater 309-311: 1235-1238.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献