The Effect of the Operational Conditions on the Quantitative Leak Rate of Leak Detectors
-
Published:2006-10
Issue:
Volume:321-323
Page:1560-1563
-
ISSN:1662-9795
-
Container-title:Key Engineering Materials
-
language:
-
Short-container-title:KEM
Author:
Shin Yong Hyeon1,
Hong Seung Soo1,
Lim In Tae2,
Chung Kwang Hwa1
Affiliation:
1. Korea Research Institute of Standards and Science
2. Korea Research Institute of Science and Standards
Abstract
The Korea Research Institute of Standards and Sciences has developed and maintained leak
standards systems, and has accumulated much data on leak standards that have been disseminated
during calibration and testing services. We have found that the indicative value of a leak detector and
a halogen leak detector varies a lot, depending on the conditions encountered during sniffer mode
operation. Therefore, to measure the reliability of the leak rate or when performing a leak test, proper
test conditions are necessary. In addition, we found that the reliability increased when we used a guide
tube that could influence the flow of air.
Publisher
Trans Tech Publications, Ltd.
Subject
Mechanical Engineering,Mechanics of Materials,General Materials Science