Affiliation:
1. National Institute of Advanced Industrial Science and Technology (AIST)
2. University of Tsukuba
3. Chiba University
Abstract
To improve the spatial resolution of positron annihilation spectroscopy (PAS), a system to produce an intense positron microbeam was developed in AIST. A slow positron beam, which was produced by an electron linear accelerator, was focused by a lens onto a remoderator to enhance its brightness. The brightness-enhanced beam with an intensity of ≈1 × 106 e+/s was extracted from the remoderator and focused onto the sample by a lens. The beam size at the sample was 25 μm, which is more than two and half orders of magnitude smaller than that in the magnetic transport system (≈10 mm). Hence, the spatial resolution of PAS with an AIST positron microbeam can be drastically improved relative to PAS using conventional methods.
Publisher
Trans Tech Publications, Ltd.
Subject
Mechanical Engineering,Mechanics of Materials,Condensed Matter Physics,General Materials Science
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献