Abstract
Background: We aimed to evaluate whether the use of our novel patient-specific guide (PSG) with 3-dimensional reconstruction in reverse total shoulder arthroplasty (RTSA) would allow accurate and reliable implantation of the glenoid and humeral components.Methods: 20 fresh-frozen cadaveric shoulders were used. The PSG group (n=10) and conventional group (n=10) was evaluated the accuracy and reproducibility of implant positioning between before and after surgery on the computed tomography image.Results: The superoinferior and anteroposterior offset in the glenoid component were 0.42 ± 0.07, 0.50 ± 0.08 in the conventional group and 0.45 ± 0.03, 0.46 ± 0.02 in the PSG group. The inclination and version angles were -1.93° ± 4.31°, 2.27° ± 5.91° and 0.46° ± 0.02°, 3.38° ± 2.79°. The standard deviation showed a smaller difference in the PSG group. The anteroposterior and lateromedial humeral canal center offset in the humeral component were 0.45 ± 0.12, 0.48 ± 0.15 in the conventional group and 0.46 ± 0.59 (<i>p</i>=0.794), 0.46 ± 0.06 (<i>p</i>=0.702) in the PSG group. The PSG showed significantly better humeral stem alignment.Conclusions: The use of PSGs with 3-dimensional reconstruction reduces variabilities in glenoid and humerus component positions and prevents extreme positioning errors in RTSA.
Funder
Institute for Information and communications Technology Promotion
Ministry of Science, ICT and Future Planning
Publisher
Korean Shoulder and Elbow Society
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献