Author:
Tomasino Andre,Gebhard Harry,Parikh Karishma,Wess Christian,Härtl Roger
Abstract
Object
The authors present the radiological and clinical outcome data obtained in patients who underwent single-level anterior cervical discectomy and fusion (ACDF) for cervical spondylosis and/or disc herniation; bioabsorbable plates were used for instrumentation. The use of metallic plates in ACDF has gained acceptance as a stabilizing part of the procedure to increase fusion rates, but when complications occur with these devices, the overall effectiveness of the procedure is compromised. As a possible solution, bioabsorbable implants for ACDF have been developed. This study investigates the feasibility and radiological and clinical outcomes of the bioabsorbable plates for ACDF.
Methods
The radiological and clinical outcomes of 30 patients were investigated retrospectively. All patients presented with cervical radiculopathy or myelopathy and underwent single-level ACDF in which a bioabsorbable anterior cervical plate and an allograft bone spacer were placed at a level between C-3 and C-7. Radiological outcome was assessed based on the fusion rate, subsidence, and Cobb angle of the surgical level. Clinical outcome was determined by using a visual analog scale, the Neck Disability Index, and the Odom criteria.
Results
There were no intraoperative complications, and no hardware failure was observed. No signs or symptoms of adverse tissue reaction caused by the implant were seen. Two reoperations were necessary due to postoperative blood collections. The overall complication rate was 16.7%. After 6 months, radiographic fusion was seen in 92.3% of patients. Subsidence at 11.3 ± 7.2 months was 3.1 ± 5.8 mm (an 8.2% change over the immediately postoperative results), and the change in the sagittal curvature was –2.7 ± 2.7°. The visual analog scale score for neck and arm pain and Neck Disability Index improved significantly after surgery (p < 0.001). Overall at 19.5 months postoperatively, 83% of the patients had favorable outcomes based on the Odom criteria.
Conclusions
Absorbable instrumentation provides better stability than the absence of a plate but graft subsidence and deformity rates may be higher than those associated with metal implants. There were no device-related complications, but adverse late effects cannot be excluded. The fusion rate and outcome are comparable to the results achieved with metallic plates. The authors were satisfied with the use of bioabsorbable plates as a reasonable alternative to metal, avoiding the need for lifelong metallic implants.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献