Bipolar radiofrequency ablation of aneurysm remnants after coil embolization can improve endovascular treatment of experimental bifurcation aneurysms

Author:

Boileau Xavier1,Zeng Han1,Fahed Robert1,Bing Fabrice12,Makoyeva Alina1,Darsaut Tim E.3,Savard Pierre4,Coutu Benoit5,Salazkin Igor1,Raymond Jean12

Affiliation:

1. Centre Hospitalier de l'Université de Montréal Research Centre (CRCHUM), Notre-Dame Hospital, Interventional Neuroradiology Research Laboratory, Montreal, Quebec;

2. Centre Hospitalier de l'Université de Montréal (CHUM), Notre-Dame Hospital, Department of Radiology, Service of Neuroradiology, Montreal, Quebec;

3. University of Alberta Hospital, Mackenzie Health Sciences Centre, Department of Surgery, Division of Neurosurgery, Edmonton, Alberta;

4. Institute of Biomedical Engineering, École Polytechnique de Montréal; and

5. Centre Hospitalier de l'Université de Montréal (CHUM), Notre-Dame Hospital, Department of Medicine, Service of Cardiology, Montreal, Quebec, Canada

Abstract

OBJECTIVEEndovascular treatment of aneurysms may result in incomplete initial occlusion and aneurysm recurrence at angiographic follow-up studies. This study aimed to assess the feasibility and efficacy of bipolar radiofrequency ablation (RFA) of aneurysm remnants after coil embolization.METHODSBipolar RFA was accomplished using the coil mass as 1 electrode, while the second electrode was a stent placed across the aneurysmal neck. After preliminary experiments and protocol approval from the Animal Care committee, wide-necked bifurcation aneurysms were constructed in 24 animals. Aneurysms were allocated to 1 of 3 groups: partial intraoperative coil embolization, followed by RFA (n = 12; treated group) or without RFA (n = 6; control group 1); or attempted complete endovascular coil embolization 2–4 weeks later (n = 6; control group 2). Angiographic results were compared at baseline, immediately after RFA, and at 12 weeks, using an ordinal scale. Pathological results and neointima formation at the neck were compared using a semiquantitative grading scale.RESULTSBipolar RFA was able to reliably target the aneurysm neck when the coil mass and stent were used as electrodes. RFA improved angiographic results immediately after partial coiling (p = 0.0024). Two RFA-related complications occurred, involving transient occlusion of 1 carotid artery and 1 hemorrhage from an adventitial arterial blister. At 12 weeks, angiographic results were improved with RFA (median score of 0), when compared with controls (median score of 2; p = 0.0013). Neointimal closure of the aneurysm neck was better with RFA compared with controls (p = 0.0003).CONCLUSIONSBipolar RFA can improve results of embolization in experimental models by selectively ablating residual lesions after coil embolization.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3