Affiliation:
1. Hofstra Northwell School of Medicine, Hempstead; and
2. Departments of Neurosurgery and
3. Radiology, North Shore University Hospital, Manhasset, New York
Abstract
OBJECTIVE
Complications of laser interstitial thermal therapy (LITT) are underreported. The authors discuss how they have modified their technique in the context of technical and treatment-related adverse events.
METHODS
The Medtronic Visualase system was used in 49 procedures in 46 patients. Between 1 and 3 cooling catheters/laser fiber assemblies were placed, for a total of 62 implanted devices. Devices were placed using frameless stereotaxy (n = 3), frameless stereotaxy with intraoperative MRI (iMRI) (n = 9), iMRI under direct vision (n = 2), MRI alone (n = 1), or frame-based (n = 47) techniques. LITT was performed while monitoring MRI thermometry. Indications included brain tumors (n = 12), radiation necrosis (n = 2), filum terminale ependymoma (n = 1), mesial temporal lobe epilepsy (n = 21), corpus callosotomy for bifrontal epilepsy (n = 3), cavernoma (n = 1), and hypothalamic hamartomas (n = 6).
RESULTS
Some form of adverse event occurred in 11 (22.4%) of 49 procedures. These included 4 catheter malpositions, 3 intracranial hemorrhages, 3 cases of neurological deficit related to thermal injury, and 1 technical malfunction resulting in an aborted procedure. Of these, direct thermal injury was the only cause of prolonged neurological morbidity and occurred in 3 of 49 procedures. Use of frameless stereotaxy and increased numbers of devices were associated with significantly increased complication rates (p < 0.05). A number of procedural modifications were made to avoid complications, including the use of 1) frame-based catheter placement, a 1.8-mm alignment rod to create a track and titanium skull anchors for long trajectories to improve accuracy; 2) a narrow-gauge instrument for dural puncture and coregistration of contrast MRI with CT angiography to reduce intracranial hemorrhage; 3) general endotracheal anesthesia for posterior-placed skull anchors to reduce the likelihood of damage to the cooling catheter; 4) use of as few probes as possible to reduce complications overall; and 5) dose modification of thermal treatment and use of short (3-mm) diffusing tips to limit treatment when structures to be spared do not have intervening CSF spaces to act as heat sinks.
CONCLUSIONS
Laser ablation treatment may be used for a variety of neurosurgical procedures for patients with tumors and epilepsy. While catheter placement and thermal treatment may be associated with a range of suboptimal operative and postoperative courses, permanent neurological morbidity is less common. The authors' institutional experience illustrates a number of measures that may be taken to improve outcomes using this important new tool in the neurosurgical arsenal.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Subject
Genetics,Animal Science and Zoology
Cited by
118 articles.