A robotic MR-guided high-intensity focused ultrasound platform for intraventricular hemorrhage: assessment of clot lysis efficacy in a brain phantom

Author:

Raghuram Hrishikesh12,Looi Thomas13,Pichardo Samuel45,Waspe Adam C.16,Drake James M.1237

Affiliation:

1. Posluns Centre for Image Guided Innovation and Therapeutic Intervention, Hospital for Sick Children, Toronto, Ontario;

2. The Institute of Biomedical Engineering, University of Toronto, Ontario;

3. Mechanical Engineering, and

4. Radiology and Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Alberta; and

5. Hotchkiss Brain Institute, University of Calgary, Alberta, Canada

6. Departments of Medical Imaging,

7. Neurosurgery, University of Toronto, Ontario;

Abstract

OBJECTIVE Intraventricular hemorrhage (IVH) is a neurovascular complication due to premature birth that results in blood clots forming within the ventricles. Magnetic resonance–guided high-intensity focused ultrasound (MRgHIFU) has been investigated as a noninvasive treatment to lyse clots. The authors designed and constructed a robotic MRgHIFU platform to treat the neonatal brain that facilitates ergonomic patient positioning. The clot lysis efficacy of the platform is quantified using a brain phantom and clinical MRI system. METHODS A thermosensitive brain-mimicking phantom with ventricular cavities was developed to test the clot lysis efficacy of the robotic MRgHIFU platform. Whole porcine blood was clotted within the phantom’s cavities. Using the MRgHIFU platform and a boiling histotripsy treatment procedure (500 W, 10-msec pulse duration, 1.0% duty cycle, and 40-second duration), the clots were lysed inside the phantom. The contents of the cavities were vacuum filtered, and the remaining mass of the solid clot particles was used to quantify the percentage of clot lysis. The interior of the phantom’s cavities was inspected for any collateral damage during treatment. RESULTS A total of 9 phantoms were sonicated, yielding an average (± SD) clot lysis of 97.0% ± 2.57%. Treatment resulted in substantial clot lysis within the brain-mimicking phantoms that were apparent on postsonication T2-weighted MR images. No apparent collateral damage was observed within the phantom after treatment. The results from the study showed the MRgHIFU platform was successful at lysing more than 90% of a blood clot at a statistically significant level. CONCLUSIONS The robotic MRgHIFU platform was shown to lyse a large percentage of a blood clot with no observable collateral damage. These results demonstrate the platform’s ability to induce clot lysis when targeting through simulated brain matter and show promise toward the final application in neonatal patients.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Body-mounted MR-HIFU Robotic System: Mechanical Design and Accuracy Evaluation;2023 57th Annual Conference on Information Sciences and Systems (CISS);2023-03-22

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3