Programming jammed Codman Hakim programmable valves: study of an explanted valve and successful programming in a patient

Author:

Wong Sui-To,Wen Eleanor,Fong Dawson

Abstract

Malfunction of a Codman Hakim programmable valve due to jamming of its programmable component may necessitate shunt revision. The authors report a method for programming jammed Codman Hakim programmable valves by using a Strata II magnet and additional neodymium magnets. The programming method was derived after studying a jammed valve in the laboratory that was explanted from an 10-year-old boy with a history of fourth ventricle ependymoma. Programming the explanted valve with a Codman programmer failed, but rotating a Strata II magnet above the valve resulted in rotation of the spiral cam in the valve. It was found that the Strata II magnet could be used to program the jammed valve by rotating the magnet 90° or multiples of 90° above the valve. The strength of the magnetic field of the Strata II magnet was able to be increased by putting neodymium magnets on it. The programming method was then successfully used in a patient with a jammed Codman Hakim programmable valve. After successful programming using this method, clinical and radiological follow-up of the patient was advised.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3