Abstract
The authors report an unusual case of intrathecal baclofen withdrawal due to the perforation and subsequent leakage of a baclofen pump catheter in a patient with spastic cerebral palsy. A 15-year-old boy underwent an uncomplicated placement of an intrathecal baclofen pump for the treatment of spasticity due to cerebral palsy. After excellent control of symptoms for 3 years, the patient presented to the emergency department with increasing tremors following a refill of his baclofen pump. Initial evaluation consisted of radiographs of the pump and catheter, which appeared normal, and a successful aspiration of CSF from the pump’s side port. A CT dye study revealed a portion of the catheter directly overlying the refill port and extravasation of radiopaque dye into the subfascial pocket anterior to the pump. During subsequent revision surgery, a small puncture hole in the catheter was seen to be leaking the drug. The likely cause of the puncture was an inadvertent perforation of the catheter by a needle during the refilling of the pump. This case report highlights a unique complication in a patient with an intrathecal baclofen pump. Physicians caring for these patients should be aware of this rare yet potential complication in patients presenting with baclofen withdrawal symptoms.