Overheated and melted intracranial pressure transducer as cause of thermal brain injury during magnetic resonance imaging

Author:

Tanaka Reiichiro1,Yumoto Tetsuya2,Shiba Naoki1,Okawa Motohisa3,Yasuhara Takao4,Ichikawa Tomotsugu5,Tokunaga Koji5,Date Isao5,Ujike Yoshihito1

Affiliation:

1. Department of Critical Care and Emergency Medicine, Okayama University Hospital;

2. Department of Surgery, National Hospital Organization Mito Medical Center, Ibaraki; and

3. Department of Health Science, College of Life Science, Kurashiki University of Science and The Arts;

4. Spinal Cord Treatment Center, Shin Komonji Hospital, Fukuoka, Japan

5. Department of Neurological Surgery, Okayama University Graduate School of Medicine, Okayama;

Abstract

Magnetic resonance imaging is used with increasing frequency to provide accurate clinical information in cases of acute brain injury, and it is important to ensure that intracranial pressure (ICP) monitoring devices are both safe and accurate inside the MRI suite. A rare case of thermal brain injury during MRI associated with an overheated ICP transducer is reported. This 20-year-old man had sustained a severe contusion of the right temporal and parietal lobes during a motor vehicle accident. An MR-compatible ICP transducer was placed in the left frontal lobe. The patient was treated with therapeutic hypothermia, barbiturate therapy, partial right temporal lobectomy, and decompressive craniectomy. Immediately after MRI examination on hospital Day 6, the ICP monitor was found to have stopped working, and the transducer was subsequently removed. The patient developed meningitis after this event, and repeat MRI revealed additional brain injury deep in the white matter on the left side, at the location of the ICP transducer. It is suspected that this new injury was caused by heating due to the radiofrequency radiation used in MRI because it was ascertained that the tip of the transducer had been melted and scorched. Scanning conditions—including configuration of the transducer, MRI parameters such as the type of radiofrequency coil, and the specific absorption rate limit—deviated from the manufacturer's recommendations. In cooperation with the manufacturer, the authors developed a precautionary tag describing guidelines for safe MR scanning to attach to the display unit of the product. Strict adherence to the manufacturer's guidelines is very important for preventing serious complications in patients with ICP monitors undergoing MRI examinations.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献