Accuracy of pedicle screw insertion in the cervical spine for internal fixation using frameless stereotactic guidance

Author:

Rath Stefan A.12,Moszko Slawomir1,Schäffner Petra M.1,Cantone Giuseppe1,Braun Veit2,Richter Hans-Peter2,Antoniadis Gregor2

Affiliation:

1. 1Department of Neurosurgery and Interventional Neuroradiology, Medical Center, Deggendorf, Germany; and

2. 2Department of Neurosurgery, District Hospital Günzburg, University of Ulm, Günzburg, Germany

Abstract

Object Although transpedicular fixation is a biomechanically superior technique, it is not routinely used in the cervical spine. The risk of neurovascular injury in this region is considered high because the diameter of cervical pedicles is very small and their angle of insertion into the vertebral body varies. This study was conducted to analyze the clinical accuracy of stereotactically guided transpedicular screw insertion into the cervical spine. Methods Twenty-seven patients underwent posterior stabilization of the cervical spine for degenerative instability resulting from myelopathy, fracture/dislocation, tumor, rheumatoid arthritis, and pyogenic spondylitis. Fixation included 1–6 motion segments (mean 2.2 segments). Transpedicular screws (3.5-mm diameter) were placed using 1 of 2 computer-assisted guidance systems and lateral fluoroscopic control. The intraoperative mean deviation of frameless stereotaxy was < 1.9 mm for all procedures. Results No neurovascular complications resulted from screw insertion. Postoperative computed tomography (CT) scans revealed satisfactory positioning in 104 (90%) of 116 cervical pedicles and in all 12 thoracic pedicles. A noncritical lateral or inferior cortical breach was seen with 7 screws (6%). Critical malplacement (4%) was always lateral: 5 screws encroached into the vertebral artery foramen by 40–60% of its diameter; Doppler sonographic controls revealed no vascular compromise. Screw malplacement was mostly due to a small pedicle diameter that required a steep trajectory angle, which could not be achieved because of anatomical limitation in the exposure of the surgical field. Conclusions Despite the use of frameless stereotaxy, there remains some risk of critical transpedicular screw malpositioning in the subaxial cervical spine. Results may be improved by the use of intraoperative CT scanning and navigated percutaneous screw insertion, which allow optimization of the transpedicular trajectory.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3