Author:
McClelland Shearwood,Ford Blair,Senatus Patrick B.,Winfield Linda M.,Du Yunling E.,Pullman Seth L.,Yu Qiping,Frucht Steven J.,McKhann Guy M.,Goodman Robert R.
Abstract
Object
Deep brain stimulation (DBS) of the subthalamic nucleus (STN) performed using intraoperative microelectrode recording (MER) to adjust electrode placement has become a widely used treatment for patients with advanced Parkinson disease (PD). Few studies have been conducted to examine the location of implanted electrodes relative to the intended target, and even fewer have been undertaken to investigate the degree to which variations in the location of these electrodes impacts their clinical efficacy. This study was performed to examine these issues.
Methods
The authors located 52 bilaterally implanted DBS electrode tips on postoperative magnetic resonance (MR) images obtained in 26 consecutive patients. Postoperative and preoperative planning MR images were merged to determine the DBS electrode tip coordinates relative to the midcommissural point. Surgical records listed the intended target coordinates for each DBS electrode tip. Clinical outcome assessment included the Unified PD Rating Scale (UPDRS) motor score at 1 year, standardized questionnaires, and routine follow-up visits.
The mean difference between electrode tip location and intended target for all 52 electrodes was less than 2 mm in all axes. Only one electrode was farther than 3 mm from the intended target, and this was the only electrode that had to be replaced due to lack of clinical efficacy (lack of tremor suppression); its reimplantation 4 mm more medially provided excellent tremor control. High correlation coefficients indicate that the MR imaging analysis accurately determined the anatomical location of the electrode tips. Blinded videotape reviews of UPDRS motor scores comparing effects of stimulation in patients who were “on” and “off” medication identified subgroups in whom there was minimal and maximal stimulation response. Patients in these subgroups had no differences between the MR imaging–determined actual electrode tip location and its intended location. Similarly, improvements of dyskinesias and severity of symptoms encountered during the wearing-off period for the drug did not correlate with variations of electrode tip location.
Conclusions
The findings in this study lead the authors to suggest that a DBS electrode placed anywhere within a 6-mm-diameter cylinder centered at the presumed middle of the STN (based on stereotactic atlas coordinates) provides similar clinical efficacy. Future studies may be warranted to evaluate prospectively the degree to which MER modification of the anatomically and/or image-determined target improves clinical efficacy of DBS electrodes.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Subject
Neurology (clinical),General Medicine,Surgery
Cited by
78 articles.