Engineering patient-specific drill templates and bioabsorbable posterior cervical plates: a feasibility study

Author:

Ryken Timothy C.12,Kim John1,Owen Brent D.1,Christensen Gary E.3,Reinhardt Joseph M.4

Affiliation:

1. Carver College of Medicine,

2. Department of Neurosurgery, and

3. Departments of Electrical-Computer Engineering and

4. Biomedical Engineering, University of Iowa, Iowa City, Iowa

Abstract

Object The authors describe a method of using computer models to generate customized cervical implants. A promising yet challenging technique in cervical spine surgery involves the use of pedicle screws to assist with posterior instrumentation. Surrounding anatomical structures such as the vertebral arteries and cervical nerve roots present challenges for safe screw placement; however, the use of computer-generated templates seems to be a promising method to assist with placement. In this study, the authors explore the use of computer-generated templates and introduce their methods for creating custom, bioabsorbable posterior cervical implants. Methods The cervical spines (C2–T1) from 4 cadavers were scanned with volumetric CT. Using commercially available software, the authors generated volumetric models of a cervical drill template and the mold for a cervical plate spanning a desired number of vertebrae. The computer generated models of the cervical drill template and cervical plate mold were converted into physical models using a rapid prototyping machine. The biopolymer polylactic acid resin was heated to 250°C and resolidified to form thin ~ 5-mm-thick plates. The newly formed plates were reheated to 60°C and cast on the cervical mold. Results The resulting translucent plates were found on visual inspection to have a secure lock-and-key fit on the original cadaver spine, and the techniques used were robust and reproducible. The process described in this brief report provides the background to proceed with development and testing of these patient-absorbable templates. Conclusions The creation and use of patient-specific bioabsorbable posterior cervical plates in conjunction with multilevel drill templates appear promising. Additional feasibility studies are planned, and in vitro studies are required to determine the safety and efficacy of using patient-specific drill templates and converting them into bioabsorbable implants.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3