Monitoring of cerebral oxygenation in patients with severe head injuries: brain tissue PO2 versus jugular vein oxygen saturation

Author:

Kiening Karl L.,Unterberg Andreas W.,Bardt Tillman F.,Schneider Gerd-Helge,Lanksch Wolfgang R.

Abstract

✓ Monitoring of cerebral oxygenation is considered to be of great importance in minimizing secondary hypoxic and ischemic brain damage following severe head injury. Although the threshold for cerebral hypoxia in jugular bulb oximetry (measurement of O2 saturation in the jugular vein (SjvO2)) is generally accepted to be 50% oxygen saturation, a comparable value in brain tissue PO2 (PtiO2) monitoring, a new method for direct assessment of PO2 in the cerebral white matter, has not yet been established. Hence, the purpose of this study was to compare brain PtiO2 with SjvO2 in severely head injured patients during phases of reduced cerebral perfusion pressure (CPP) to define a threshold in brain PtiO2 monitoring. In addition, the safety and data quality of both SjvO2 and brain PtiO2 monitoring were studied. In 15 patients with severe head injuries, SjvO2 and brain PtiO2 were monitored simultaneously. For brain PtiO2 monitoring a polarographic microcatheter was inserted in the frontal cerebral white matter, whereas for SjvO2 measurements were obtained by using a fiberoptic catheter placed in the jugular bulb. Intracranial pressure was monitored by means of an intraparenchymal catheter. Mean arterial blood pressure, CPP, end-tidal CO2, and arterial oxygen saturation (pulse oximetry) were continuously recorded. All data were simultaneously stored and analyzed using a multimodal computer system. For specific analysis, phases of marked deterioration in systemic blood pressure and consecutive reductions in CPP were investigated. There were no complications that could be attributed to the PtiO2 catheters, that is, no intracranial bleeding or infection. The “time of good data quality” was 95% in brain PtiO2 compared to 43% in SjvO2; PtiO2 monitoring could be performed twice as long as SjvO2 monitoring. During marked decreases in CPP, SjvO2 and brain PtiO2 correlated closely. A significant second-order regression curve of SjvO2 versus brain PtiO2 (p < 0.01) was plotted. At a threshold of 50% in SjvO2, brain PtiO2 was found to be within the range of 3 to 12 mm Hg, with a regression curve “best fit” value of 8.5 mm Hg. There was a close correlation between CPP and oxygenation parameters (PtiO2 and SjvO2) when CPP fell below a breakpoint of 60 mm Hg, suggesting intact cerebral autoregulation in most patients. This study demonstrates that monitoring brain PtiO2 is a safe, reliable, and sensitive diagnostic method to follow cerebral oxygenation. In comparison to SjvO2, PtiO2 is more suitable for long-term monitoring. It can be used to minimize episodes of secondary cerebral maloxygenation after severe head injury and may, hopefully, improve the outcome in severely head injured patients.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Cited by 367 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3