A novel miniature robotic device for frameless implantation of depth electrodes in refractory epilepsy

Author:

Dorfer Christian1,Minchev Georgi1,Czech Thomas1,Stefanits Harald1,Feucht Martha2,Pataraia Ekaterina3,Baumgartner Christoph4,Kronreif Gernot5,Wolfsberger Stefan1

Affiliation:

1. Department of Neurosurgery; and

2. Departments of Pediatrics and Adolescence Medicine and

3. Neurology, Epilepsy Monitoring Unit, Medical University Vienna;

4. Second Neurological Department, General Hospital Hietzing, Vienna; and

5. Austrian Center of Medical Innovation and Technology (ACMIT), Wiener Neustadt, Austria

Abstract

OBJECTIVEThe authors' group recently published a novel technique for a navigation-guided frameless stereotactic approach for the placement of depth electrodes in epilepsy patients. To improve the accuracy of the trajectory and enhance the procedural workflow, the authors implemented the iSys1 miniature robotic device in the present study into this routine.METHODSAs a first step, a preclinical phantom study was performed using a human skull model, and the accuracy and timing between 5 electrodes implanted with the manual technique and 5 with the aid of the robot were compared. After this phantom study showed an increased accuracy with robot-assisted electrode placement and confirmed the robot's ability to maintain stability despite the rotational forces and the leverage effect from drilling and screwing, patients were enrolled and analyzed for robot-assisted depth electrode placement at the authors' institution from January 2014 to December 2015. All procedures were performed with the S7 Surgical Navigation System with Synergy Cranial software and the iSys1 miniature robotic device.RESULTSNinety-three electrodes were implanted in 16 patients (median age 33 years, range 3–55 years; 9 females, 7 males). The authors saw a significant increase in accuracy compared with their manual technique, with a median deviation from the planned entry and target points of 1.3 mm (range 0.1–3.4 mm) and 1.5 mm (range 0.3–6.7 mm), respectively. For the last 5 patients (31 electrodes) of this series the authors modified their technique in placing a guide for implantation of depth electrodes (GIDE) on the bone and saw a significant further increase in the accuracy at the entry point to 1.18 ± 0.5 mm (mean ± SD) compared with 1.54 ± 0.8 mm for the first 11 patients (p = 0.021). The median length of the trajectories was 45.4 mm (range 19–102.6 mm). The mean duration of depth electrode placement from the start of trajectory alignment to fixation of the electrode was 15.7 minutes (range 8.5–26.6 minutes), which was significantly faster than with the manual technique. In 12 patients, depth electrode placement was combined with subdural electrode placement. The procedure was well tolerated in all patients. The authors did not encounter any case of hemorrhage or neurological deficit related to the electrode placement. In 1 patient with a psoriasis vulgaris, a superficial wound infection was encountered. Adequate physiological recordings were obtained from all electrodes. No additional electrodes had to be implanted because of misplacement.CONCLUSIONSThe iSys1 robotic device is a versatile and easy to use tool for frameless implantation of depth electrodes for the treatment of epilepsy. It increased the accuracy of the authors' manual technique by 60% at the entry point and over 30% at the target. It further enhanced and expedited the authors' procedural workflow.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

Reference36 articles.

1. A novel miniature robotic guidance device for stereotactic neurosurgical interventions: preliminary experience with the iSYS1 robot;Minchev;J Neurosurg

2. The application accuracy of stereotactic frames;Maciunas;Neurosurgery,1994

3. Insertion of depth electrodes with or without subdural grids using frameless stereotactic guidance systems—technique and outcome;Murphy;Br J Neurosurg,2002

4. Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation;Bjartmarz;Stereotact Funct Neurosurg,2007

5. Frameless stereotaxy using bone fiducial markers for deep brain stimulation;Holloway;J Neurosurg,2005