Abstract

OBJECTIVE The object of this study was to compare the efficacy and safety profile of the Pipeline embolization device (PED)/Pipeline Flex embolization device (PED Flex) with that of the Pipeline Flex embolization device with Shield Technology (PED Shield). After introducing the first-generation PED and the second-generation PED Flex with its updated delivery system, the PED Shield was launched with a synthetic layer of phosphorylcholine surface modification to reduce thrombogenicity. METHODS This is a retrospective review of unruptured aneurysms treated with PED/PED Flex versus PED Shield between 2017 and 2022 at the authors’ institution. Patients with ruptured aneurysms, adjunctive treatment, failed flow diverter deployment, and prior treatment of the target aneurysm were excluded. Baseline characteristics were collected for all patients, including age, sex, past medical history (hypertension, hyperlipidemia, diabetes mellitus), smoking status, aneurysm location, and aneurysm dimensions (neck, width, height) and morphology (saccular, nonsaccular). The primary outcome was procedural and periprocedural complication rates. RESULTS The study cohort comprised 200 patients with 200 aneurysms, including 150 aneurysms treated with the PED/PED Flex and 50 treated with the PED Shield. With respect to intraprocedural and periprocedural complications, length of stay, length of follow-up, and functional outcome at discharge, there was no significant difference between the two cohorts. At the midterm follow-up, the rate of in-stent stenosis (PED/PED Flex: 14.2% vs PED Shield: 14.6%, p = 0.927), aneurysm occlusion (complete occlusion: 79.5% vs 80.5%, respectively; neck remnant: 4.7% vs 12.2%; dome remnant: 15.7% vs 7.3%; p = 0.119), and the need for retreatment (5.3% vs 0%, p = 0.097) were comparable between the two cohorts. CONCLUSIONS This study suggests that, as compared to first- and second-generation PED and PED Flex, the third-generation PED Shield offers similar rates of complications, aneurysm occlusion, and in-stent stenosis at the midterm follow-up.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

Reference31 articles.

1. Current treatment strategies for intracranial aneurysms: an overview;Zhao J,2018

2. Pipeline Flex embolisation device with Shield Technology for the treatment of patients with intracranial aneurysms: periprocedural and 6 month outcomes;Yeomans J,2020

3. Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial;Becske T,2013

4. Surgical or endovascular treatment of MCA aneurysms: an agreement study;Boisseau W,2022

5. International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion;Molyneux AJ,2005

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3