Does new magnetic resonance imaging technology provide better geometrical accuracy during stereotactic imaging?

Author:

Novotny Josef,Vymazal Josef,Novotny Josef,Tlachacova Daniela,Schmitt Michal,Chuda Pavel,Urgosik Dusan,Liscak Roman

Abstract

Object.The authors sought to compare the accuracy of stereotactic target imaging using the Siemens 1T EXPERT and 1.5T SYMPHONY magnetic resonance (MR) units.Methods.A water-filled cylindrical Perspex phantom with axial and coronal inserts containing grids of glass rods was fixed in the Leksell stereotactic frame and subjected to MR imaging in Siemens 1T EXPERT and Siemens 1.5T SYMPHONY units. Identical sequences were used for each unit. The images were transferred to the GammaPlan treatment planning system. Deviations between stereotactic coordinates based on MR images and estimated real geometrical positions given by the construction of the phantom insert were evaluated for each study. The deviations were further investigated as a function of the MR unit used, MR sequence, the image orientation, and the spatial position of measured points in the investigated volume.Conclusions.Larger distortions were observed when using the SYMPHONY 1.5T unit than those with the EXPERT 1T unit. Typical average distortion in EXPERT 1T was not more than 0.6 mm and 0.9 mm for axial and coronal images, respectively. Typical mean distortion for SYMPHONY 1.5T was not more than 1 mm and 1.3 mm for axial and coronal images, respectively. The image sequence affected the distortions in both units. Coronal T2-weighted spin-echo images performed in subthalamic imaging produced the largest distortions of 2.6 mm and 3 mm in the EXPERT 1T and SYMPHONY 1.5T, respectively. Larger distortions were observed in coronal slices than in axial slices in both units, and this effect was more pronounced in SYMPHONY 1.5T. Noncentrally located slice positions in the investigated volume of the phantom were associated with larger distortions.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3