Evaluation of Strata NSC and Codman Hakim adjustable cerebrospinal fluid shunts and their corresponding antisiphon devices

Author:

Arnell Kai1,Koskinen Lars-Owe D.2,Malm Jan2,Eklund Anders3

Affiliation:

1. Department of Paediatric Surgery, University Hospital, Uppsala;

2. Department of Clinical Neuroscience, Umeå University, Umeå; and

3. Biomedical Engineering and Informatics, Umeå University Hospital, Umeå, Sweden

Abstract

Object The authors investigated and compared the in vitro characteristics of 2 CSF shunts, the Strata NSC and the Codman Hakim, and their corresponding antisiphon devices (ASDs). Methods Six new CSF shunts and the corresponding ASDs for each model were tested in an automated, computerized experimental setup based on pressure regulation. Opening pressure accuracy, resistance, sensitivity to abdominal pressure, antisiphon effect, and the influence of different ASD positions were determined. Results In general the shunts performed according to the manufacturers' specifications. However, at the lowest setting, the opening pressure of the Strata NSC was close to 0, and in the Codman Hakim shunt, it was higher than specified. The resistance in the Codman Hakim shunt (5.4 mm Hg/ml/min) was much higher than that in the Strata NSC (3.6 mm Hg/ml/min). Abdominal pressure affected opening pressure in both valves. Positioning the Strata ASD above or below the ventricular catheter tip resulted in higher and lower opening pressures, respectively, than when it was placed in line with the catheter. The positioning of the Codman Hakim ASD did not influence the opening pressure. Conclusions Both CSF shunts work properly, but at the lowest setting the opening pressure of the Strata NSC was near 0 and in the Codman Hakim it was twice the manufacturer's specifications. The resistance in the Strata NSC was below the normal physiological range, and in the Codman Hakim device it was in the lower range of normal. The ASD did not change the shunt characteristics in the lying position and therefore might not do so in children. If this is the case, then a shunt system with an integrated ASD could be implanted at the first shunt insertion, thus avoiding a second operation and the possibility of infection.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3