Author:
Morshed Ramin A.,Han Seunggu J.,Lau Darryl,Berger Mitchel S.
Abstract
Surgery guided by 5-aminolevulinic acid (ALA) fluorescence has become a valuable adjunct in the resection of malignant intracranial gliomas. Furthermore, the fluorescence intensity of biopsied areas of a resection cavity correlates with histological identification of tumor cells. However, in the case of lesions deep within a resection cavity, light penetration may be suboptimal, resulting in less excitation of 5-ALA metabolites, leading to decreased fluorescence emission. To address this obstacle, the authors report on the use of a 400-nm wavelength fiber-optic lighted suction instrument that can be used both during resection of a tumor and to provide direct light to deeper areas of a resection cavity. In the presented case, this wavelength-specific lighted suction instrument improved the fluorescence intensity of patches of malignant tissue within the resection cavity. This technique may further improve the utility of 5-ALA in identifying tumor-infiltrated tissue for deep-seated lesions. Additionally, this tool may have implications for scoring systems that correlate 5-ALA fluorescence intensity with histological identification of malignant cells.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Subject
Genetics,Animal Science and Zoology
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献