Intraoperative auditory steady-state monitoring during surgery in the cerebellopontine angle for estimation of postoperative hearing classes

Author:

Rampp Stefan1,Rensch Leonhard1,Simmermacher Sebastian1,Rahne Torsten2,Strauss Christian1,Prell Julian1

Affiliation:

1. Departments of Neurosurgery and

2. Otorhinolaryngology, Head and Neck Surgery, University Hospital Halle (Saale), Germany

Abstract

OBJECTIVEBrainstem auditory evoked potentials (BAEPs) have been used for intraoperative monitoring of the auditory nerve for many years. However, BAEPs yield limited information about the expected postoperative hearing quality and speech perception. The auditory steady-state response (ASSR) enables objective audiograms to be obtained in patients under anesthesia. These ASSRs could be used for intraoperative estimation of hearing classes to improve the postoperative outcome and quality of life. Studies investigating the clinical use of ASSRs during total intravenous anesthesia are currently lacking. The work presented in this article therefore investigates the application of ASSRs for intraoperative estimation of hearing classes.METHODSIn 43 patients undergoing surgery for vestibular schwannoma, ASSR measurements were performed at the beginning and end of the surgical procedure. ASSR stimuli consisted of 80-dB hearing level amplitude-modulated tones with 5-minute duration, 90-Hz modulation, and 3 different carrier frequencies: 500, 1000, and 2000 Hz. Stimulation was performed unilaterally with and without contralateral masking, using single and combined carriers. Evoked responses were recorded and analyzed in the frequency domain. ASSRs were compared with extraoperative hearing classes and BAEPs using ANOVA, correlation, and receiver operating characteristic statistics.RESULTSASSRs yielded high and consistent area under the curve (AUC) values (mean 0.83) and correlation values (mean −0.63), indicating reliable prediction of hearing classes. Analysis of BAEP amplitude changes showed lower AUC (mean 0.79) and correlation values (0.63, 0.37, and 0.50 for Waves I, III, and V, respectively). Latencies showed low AUC values (mean 0.6) and no significant correlation. Combination of several carriers for simultaneous evaluation reduced ASSR amplitudes and respective AUC values. Contralateral masking did not show a significant effect.CONCLUSIONSASSRs robustly estimate hearing class in patients under total intravenous anesthesia, even when using short measurement durations. The method provides a diagnostic performance that exceeds conventional BAEP monitoring and enables objective and automated evaluation. On the basis of these findings, continuous intraoperative auditory monitoring could become a promising alternative or adjunct to BAEPs.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3